Adjustment of the evoked response sensitivity after hospital discharge in pacemaker patients with automatic ventricular threshold tracking activated.

A Schuchert1, R Ventura, T Meinertz.   

Abstract

Automatic threshold tracking in cardiac pacemakers allows ventricular capture verification and self-adaptive pacing output regulation. The Autocapture algorithm detects the evoked response (ER) signal immediately after the pacing pulse to verify the efficacy of ventricular pacing. Before hospital delivery, the ER sensitivity must be programmed individually so that the pacemaker detects the ER signal adequately without sensing lead polarization. The aims of the study were to assess the frequency of patients in whom Autocapture could be activated and whether the ER sensitivity had to be adjusted after hospital discharge. The study included 44 patients who received the VVIR pacemaker Regency SR+ (St. Jude Medical) connected to the model 1450 T pacing lead. ER signal, lead polarization, and ER sensitivity were evaluated before hospital discharge and 1, 3, and 6 months after implantation. The system recommended activating Autocapture in 42 of 44 patients. The mean ER signal was 8.4+/-1.2 mV at discharge, 9.0+/-3.9 mV at month 1, 8.9+/-4.9 mV at month 3, and 9.3+/-4.5 mV at month 6. Polarization was 1.0+/-0.1 mV at discharge, 1.1+/-0.5 mV at month 1, 1.1+/-0.2 mV at month 3, and 1.1+/-0.5 mV at month 6. Mean ER sensitivity was 3.7+/-1.8 mV at discharge, 4.0+/-1.8 mV after 1, 4.1+/-2.2 mV after 3, and 4.1+/-1.8 mV after 6 months. ER sensitivity could remain unadjusted in 14 patients. Programming to a less sensitive ER setting from 2.9+/-1.2 mV to 4.3+/-1.5 mV was possible in 21 patients. Programming to a more sensitive ER setting from 4.1+/-1.1 mV to 2.5+/-0.9 mV was required in nine patients because of the decrease of the ER signal. The automatic threshold tracking algorithm Autocapture could be activated in 95% of patients. Programming to more sensitive ER settings was recommended in 21% of the patients after hospital discharge. Therefore, ER signal and polarization must be checked at each follow-up, as a decrease in ER signal amplitude can make reprogramming of the ER sensitivity necessary. There is no risk for the patient if the ER is not sensed, as high voltage backup stimulation is present.

Entities:  

Mesh:

Year:  2001        PMID: 11270702     DOI: 10.1046/j.1460-9592.2001.00212.x

Source DB:  PubMed          Journal:  Pacing Clin Electrophysiol        ISSN: 0147-8389            Impact factor:   1.976


  2 in total

1.  Impact of automatic adjustment of stimulation outputs on pacemaker longevity in a new dual-chamber pacing system.

Authors:  Christiane Brockes; Mariette Rahn-Schönbeck; Firat Duru; Reto Candinas; Marko Turina
Journal:  J Interv Card Electrophysiol       Date:  2003-02       Impact factor: 1.900

2.  Hemodynamic Surveillance of Ventricular Pacing Effectiveness with the Transvalvular Impedance Sensor.

Authors:  Valeria Calvi; Giovanni Pizzimenti; Marco Lisi; Giuseppe Doria; Ludovico Vasquez; Francesco Lisi; Salvatore Felis; Donatella Tempio; Alfredo Virgilio; Alberto Barbetta; Franco Di Gregorio
Journal:  Adv Med       Date:  2014-08-04
  2 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.