Biopsy of lung nodules with use of I-I device under intermittent CT fluoroscopic guidance: preliminary clinical study.

T Irie1, M Kajitani, K Matsueda, Y Arai, Y Inaba, Y Kujiraoka, Y Itai.   

Abstract

PURPOSE: To investigate the efficacy of computed tomography (CT) fluoroscopy and a new needle holder (the I-I device) in lung nodule biopsy.
MATERIALS AND METHODS: The I-I device is made of acrylate resin and was used to keep the entire needle in the tomographic plane. This study consisted of biopsies of 79 lung nodules in 77 patients. The final diagnoses were malignant in 54 patients, benign in 23, and unconfirmed in two. The biopsy procedure time from the beginning of the CT fluoroscopy procedure to the removal of the needle was measured for 24 needle passes. The radiation dose on the physician's hand was measured in five cases with use of a thermoluminescence ring.
RESULTS: Fifty-one malignant and 20 benign lesions were correctly diagnosed with histologic specimens (90%). In 58 of 77 patients (75%), the biopsy procedures were completed within a single breath-hold. Pneumothorax occurred in 20 of 77 patients (26%) and chest tube insertion was required in five. The incidence of pneumothorax was significantly lower in patients who held their breath during biopsy procedures compared with those who did not (P < .0001; chi2 test). The biopsy procedure time ranged from 15 to 39 seconds (mean: 28.2 sec). The mean radiation dose on the physician's hand was 2 mSv/case.
CONCLUSION: The diagnostic accuracy of biopsy with use of the I-I device under CT fluoroscopic guidance is comparable with that of the conventional method; however, a combination of CT fluoroscopy and the I-I device enables rapid biopsy procedures.

Entities:  

Mesh:

Year:  2001        PMID: 11265886     DOI: 10.1016/s1051-0443(07)61828-4

Source DB:  PubMed          Journal:  J Vasc Interv Radiol        ISSN: 1051-0443            Impact factor:   3.464


  9 in total

1.  Newly developed application system for CT-guided puncture based on multidetector row CT without CT fluoroscopy.

Authors:  Toshiyuki Irie; Masashi Kuramochi
Journal:  Radiat Med       Date:  2007-10-26

2.  Comparison of fully automated and semi-automated biopsy needles for lung biopsy under CT fluoroscopic guidance.

Authors:  R Yoshimatsu; T Yamagami; O Tanaka; H Miura; T Tanaka; T Suzuki; T Nishimura
Journal:  Br J Radiol       Date:  2011-08-09       Impact factor: 3.039

3.  Transthoracic needle biopsy using a C-arm cone-beam CT system: diagnostic accuracy and safety.

Authors:  M J Choi; Y Kim; Y S Hong; S S Shim; S M Lim; J K Lee
Journal:  Br J Radiol       Date:  2011-07-26       Impact factor: 3.039

4.  Clinical value of CT-guided needle biopsy for retroperitoneal lesions.

Authors:  Yuki Tomozawa; Yoshitaka Inaba; Hidekazu Yamaura; Yozo Sato; Mina Kato; Takaaki Kanamoto; Makoto Sakane
Journal:  Korean J Radiol       Date:  2011-04-26       Impact factor: 3.500

5.  Dose to the interventional radiologist in CTF-guided procedures.

Authors:  J G Alves; S Sarmento; J S Pereira; M F Pereira; M J Sousa; L Cunha; A Dias; A D Oliveira; J V Cardoso; L M Santos; J Lencart; M Gouvêa; J A M Santos
Journal:  Radiat Environ Biophys       Date:  2019-04-16       Impact factor: 1.925

6.  Diagnostic performance of percutaneous lung biopsy using automated biopsy needles under CT-fluoroscopic guidance for ground-glass opacity lesions.

Authors:  T Yamagami; R Yoshimatsu; H Miura; K Yamada; A Takahata; T Matsumoto; T Hasebe
Journal:  Br J Radiol       Date:  2013-02       Impact factor: 3.039

7.  Combined fluoroscopy- and CT-guided transthoracic needle biopsy using a C-arm cone-beam CT system: comparison with fluoroscopy-guided biopsy.

Authors:  Joo Yeon Cheung; Yookyung Kim; Sung Shine Shim; Soo Mee Lim
Journal:  Korean J Radiol       Date:  2011-01-03       Impact factor: 3.500

8.  Gafchromic XR-QA2 film as a complementary dosimeter for hand-monitoring in CTF-guided biopsies.

Authors:  Sandra Sarmento; Joana Pereira; Maria José Sousa; Luís Cunha; Anabela G Dias; Miguel F Pereira; Augusto D Oliveira; João V Cardoso; Luis M Santos; Margarida Gouvêa; Joana Lencart; João G Alves; João A M Santos
Journal:  J Appl Clin Med Phys       Date:  2016-01-08       Impact factor: 2.102

9.  The use of needle holders in CTF guided biopsies as a dose reduction tool.

Authors:  Sandra Sarmento; Joana S Pereira; Maria José Sousa; Luís T Cunha; Anabela G Dias; Miguel F Pereira; Augusto D Oliveira; João V Cardoso; Luís M Santos; João A M Santos; João G Alves
Journal:  J Appl Clin Med Phys       Date:  2017-11-29       Impact factor: 2.102

  9 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.