Effect of cutting flute design on cortical bone screw insertion torque and pullout strength.

S Yerby1, C C Scott, N J Evans, K L Messing, D R Carter.   

Abstract

OBJECTIVE: To determine the effect of the number and length of cutting flutes on the insertion torque and pullout strength for self-tapping 4.5-millimeter cortical bone screws.
DESIGN: Screws were self-tapped in the diaphysis of human cadaver femurs. Each of the six screw types studied had different designs with varying cutting flute lengths and numbers. Bone mineral density, insertion torque, and pullout strength were measured.
SETTING: The study was conducted at an experimental biomechanics laboratory associated with a university medical center. OUTCOME MEASUREMENTS: Insertion torque and pullout strength were normalized by the local bone mineral density.
RESULTS: The mean normalized insertion torque of the design with four full-length cutting flutes was less than the design with three full-length flutes and the two designs with one-third length flutes (p < 0.05). The mean normalized pullout strength of the screw with four full-length flutes was significantly greater than that of all screws with fewer than three flutes (p < 0.05).
CONCLUSIONS: Priorities for a cutting flute design should ideally include ease of screw insertion, minimal soft tissue irritation, and maximal screw holding power. Screws with more than two flutes were easier to insert and did not cause cortical damage during insertion. The screw with four full-length flutes showed a trend toward being the easiest to insert and having the greatest holding strength.

Entities:  

Mesh:

Year:  2001        PMID: 11265014     DOI: 10.1097/00005131-200103000-00012

Source DB:  PubMed          Journal:  J Orthop Trauma        ISSN: 0890-5339            Impact factor:   2.512


  4 in total

1.  Template-guided vs. non-guided drilling in site preparation of dental implants.

Authors:  Uta Scherer; Marcus Stoetzer; Martin Ruecker; Nils-Claudius Gellrich; Constantin von See
Journal:  Clin Oral Investig       Date:  2014-10-30       Impact factor: 3.573

2.  A surrogate long-bone model with osteoporotic material properties for biomechanical testing of fracture implants.

Authors:  Mark B Sommers; Daniel C Fitzpatrick; Steven M Madey; Corey Vande Zanderschulp; Michael Bottlang
Journal:  J Biomech       Date:  2007-06-18       Impact factor: 2.712

3.  Drilling Capability of Orthodontic Miniscrews: In Vitro Study.

Authors:  Alessandra Marchi; Matteo Camporesi; Maurizio Festa; Luis Salvatierra; Sara Izadi; Giampietro Farronato
Journal:  Dent J (Basel)       Date:  2020-12-21

4.  Comparison and prediction of pullout strength of conical and cylindrical pedicle screws within synthetic bone.

Authors:  Wen-Chi Tsai; Po-Quang Chen; Tung-Wu Lu; Shing-Sheng Wu; Kao-Shang Shih; Shang-Chih Lin
Journal:  BMC Musculoskelet Disord       Date:  2009-04-30       Impact factor: 2.362

  4 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.