Literature DB >> 11238293

Standardization of cardiac troponin I assays: round Robin of ten candidate reference materials.

R H Christenson1, S H Duh, F S Apple, G S Bodor, D M Bunk, J Dalluge, M Panteghini, J D Potter, M J Welch, A H Wu, S E Kahn.   

Abstract

BACKGROUND: Cardiac troponin I (cTnI) results vary 100-fold among assays. As a step toward standardization, we examined the performance of 10 candidate reference materials (cRMs) in dilution studies with 13 cTnI measurement systems.
METHODS: Solutions of 10 cTnI cRMs, each characterized by NIST, were shipped to the manufacturers of 13 cTnI measurement systems. Manufacturers used their respective diluents to prepare each cRM in cTnI concentrations of 1, 10, 25, and 50 microg/L. For the purpose of ranking the cRMs, the deviation of each cTnI measurement from the expected response was assessed after normalization with the 10 microg/L cTnI solution. Normalized deviations were examined in five formats. Parameters from linear regression analysis of the measured cTnI vs expected values were also used to rank performance of the cRMs.
RESULTS: The three cRMs demonstrating the best overall rankings were complexes of troponins C, I, and T. The matrices for these three cRMs values differed; one was reconstituted directly from the lyophilized form submitted by the supplier; one was submitted in liquid form, lyophilized at NIST, and subsequently reconstituted; and the third was evaluated in the liquid form received from the supplier. The cRM demonstrating the fourth best performance was a binary complex of troponins C and I supplied in lyophilized form and reconstituted before distribution.
CONCLUSIONS: The cRMs demonstrating the best performance characteristics in 13 cTnI analytical systems will be included in subsequent activities of the cTnI Standardization Committee of the AACC.

Entities:  

Mesh:

Substances:

Year:  2001        PMID: 11238293

Source DB:  PubMed          Journal:  Clin Chem        ISSN: 0009-9147            Impact factor:   8.327


  7 in total

1.  Reference materials and reference measurement procedures: an overview from a national metrology institute.

Authors:  David M Bunk
Journal:  Clin Biochem Rev       Date:  2007-11

2.  Quantification of cardiac troponin I in human plasma by immunoaffinity enrichment and targeted mass spectrometry.

Authors:  Nicole A Schneck; Karen W Phinney; Sang Bok Lee; Mark S Lowenthal
Journal:  Anal Bioanal Chem       Date:  2018-03-01       Impact factor: 4.142

3.  Value of N-terminal pro brain natriuretic peptide in predicting prognosis and severity of coronary artery disease in acute coronary syndrome.

Authors:  Hanan Radwan; Abdelhakem Selem; Kamel Ghazal
Journal:  J Saudi Heart Assoc       Date:  2014-05-05

4.  Developing multiplexed assays for troponin I and interleukin-33 in plasma by peptide immunoaffinity enrichment and targeted mass spectrometry.

Authors:  Eric Kuhn; Terri Addona; Hasmik Keshishian; Michael Burgess; D R Mani; Richard T Lee; Marc S Sabatine; Robert E Gerszten; Steven A Carr
Journal:  Clin Chem       Date:  2009-04-16       Impact factor: 8.327

Review 5.  Role of biomarkers in risk stratification of acute coronary syndrome.

Authors:  C M Nagesh; Ambuj Roy
Journal:  Indian J Med Res       Date:  2010-11       Impact factor: 2.375

6.  Commutability of possible external quality assessment materials for cardiac troponin measurement.

Authors:  Shunli Zhang; Jie Zeng; Chuanbao Zhang; Yilong Li; Haijian Zhao; Fei Cheng; Songlin Yu; Mo Wang; Wenxiang Chen
Journal:  PLoS One       Date:  2014-07-07       Impact factor: 3.240

7.  Biochemical Markers of Myocardial Damage.

Authors:  Geza S Bodor
Journal:  EJIFCC       Date:  2016-04-20
  7 in total