Threshold contrast detail detectability measurement of the fluoroscopic image quality of a dynamic solid-state digital x-ray image detector.

A G Davies1, A R Cowen, S M Kengyelics, R F Bury, T J Bruijns.   

Abstract

Solid-state digital x-ray imaging detectors of flat-panel construction will play an increasingly important role in future medical imaging facilities. Solid-state detectors that will support both dynamic (including fluoroscopic) and radiographic image recording are under active development. The image quality of an experimental solid-state digital x-ray image detector operating in a continuous fluoroscopy mode has been investigated. The threshold contrast detail detectability (TCDD) technique was used to compare the fluoroscopic imaging performance of an experimental dynamic solid-state digital x-ray image detector with that of a reference image intensifier television (IITV) fluoroscopy system. The reference system incorporated Plumbicon TV. Results were presented as a threshold detection index, or H(T)(A), curves. Measurements were made over a range of mean entrance air kerma (EAK) rates typically used in conventional IITV fluoroscopy. At the upper and mid EAK rate range (440 and 220 nGy/s) the solid-state detector outperformed the reference IITV fluoroscopy system as measured by TCDD performance. At the lowest measured EAK rate (104 nGy/s), the solid-state detector produces slightly inferior TCDD performance compared with the reference system. Although not statistically significant at this EAK rate, the difference will increase as EAK is lowered further. Overall the TCDD results and early clinical experiences support the proposition that a current design of dynamic solid-state detector produces image quality competitive with that of modern IITV fluoroscopy systems. These findings encourage the development of compact and versatile universal x-ray imaging systems based upon solid-state detector technology to support R & F and vascular/interventional applications.

Entities:  

Mesh:

Year:  2001        PMID: 11213916     DOI: 10.1118/1.1312814

Source DB:  PubMed          Journal:  Med Phys        ISSN: 0094-2405            Impact factor:   4.071


  2 in total

1.  Do flat detector cardiac X-ray systems convey advantages over image-intensifier-based systems? Study comparing X-ray dose and image quality.

Authors:  Andrew G Davies; Arnold R Cowen; Stephen M Kengyelics; Janet Moore; Mohan U Sivananthan
Journal:  Eur Radiol       Date:  2006-11-18       Impact factor: 5.315

Review 2.  Flat-panel detectors: how much better are they?

Authors:  J Anthony Seibert
Journal:  Pediatr Radiol       Date:  2006-09
  2 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.