Literature DB >> 11212381

Assessing the image quality of a CCD-based digital intraoral radiography system: application of perceptibility curve test.

Y Hayakawa1, H Kitagawa, M Wakoh, K Kuroyanagi, U Welander.   

Abstract

The perceptibility curve test is a method for evaluating the psychophysical properties of radiographic systems. The concept of the perceptibility curve is based on the minimum perceptible exposure differences that are recorded by a particular imaging system. The perceptibility curve test was applied to a Charge-Coupled Device (CCD)-based digital intraoral radiography system in this study. A test object was made of a square aluminum block, 28 mm in both height and length and 10 mm in thickness. The test object had ten contrast details in the form of round holes with diameters of 1.5 mm. The depths of the holes ranged from 0.05 mm to 0.5 mm in steps of 0.05 +/- 0.01 mm. An X-ray unit was used that operated in the range of 60 to 90 kVp. Test radiographs were shown to observers in random order. Six observers were asked to rate them regarding the number of spots representing holes that they could perceive. The smallest perceptible difference in gray level on the computer monitor, a 15-inch cathode-ray tube (CRT), was then found. The minimum difference in gray level between a perceptible detail and the background was determined. Dose response functions were employed to determine exposures corresponding to these gray levels. For each case, delta logE as a function of the number of object details was calculated. Finally the reciprocal of all values of (delta logE)min were calculated. The number of details which the observers perceived decreased linearly with decreasing delta logE values. The maximum number of details that observers perceived was eight or nine. The perceptibility curves showed that their peaks shifted slightly with a change of the kVp but that their shapes were similar at different kVp settings.

Entities:  

Mesh:

Year:  2000        PMID: 11212381     DOI: 10.2209/tdcpublication.41.9

Source DB:  PubMed          Journal:  Bull Tokyo Dent Coll        ISSN: 0040-8891


  4 in total

1.  Effect of tube potential and image receptor on the detection of natural proximal caries in primary teeth.

Authors:  Elif Sogur; B Güniz Baksı; Kaan Orhan; S Candan Paksoy; Salih Dogan; Yılmaz S Erdal; Ali Mert
Journal:  Clin Oral Investig       Date:  2010-09-14       Impact factor: 3.573

2.  Evaluation of low-contrast perceptibility in dental restorative materials under the influence of ambient light conditions.

Authors:  A D Cruz; I C Lobo; A L B Lemos; M F Aguiar
Journal:  Dentomaxillofac Radiol       Date:  2015-01-28       Impact factor: 2.419

3.  A mathematical model relating changes of grey values to changes of thicknesses of a stepwedge.

Authors:  H Geha; B Bechara; T Faddoul; M Noujeim
Journal:  Dentomaxillofac Radiol       Date:  2012-07-27       Impact factor: 2.419

4.  Evaluation of a Mathematical Model for Digital Image Enhancement.

Authors:  Hassem Geha; Ibrahim Nasseh; Marcel Noujeim
Journal:  Open Dent J       Date:  2015-07-31