Wear-through of a modular polyethylene liner: four case reports.

C A Engh1, R H Hopper, C A Engh1, J P McAuley.   

Abstract

Four patients in whom the S-ROM Total Hip System Polydial polyethylene liner was used illustrate the importance of, and difficulties in, detecting polyethylene wear-through before a complete acetabular revision is necessary. The patients in these cases had full thickness polyethylene wear-through 11 or more years after total hip arthroplasties. Dorr and Livermore wear measurements from radiographs of the patients underestimated the actual wear and did not indicate wear-through. Because excessive head penetration was not evident radiographically with the S-ROM components, the cases showed the necessity of using additional measures to diagnose wear-through, particularly with these components. Repeat radiographs at a higher kilovoltage were used to better visualize the outline of the femoral head, and hip aspiration was used to diagnose metallosis. These cases also showed the importance of yearly followup evaluations, especially 10 or more years after surgery for patients with the S-ROM Polydial locking mechanism.

Entities:  

Mesh:

Substances:

Year:  2001        PMID: 11210951     DOI: 10.1097/00003086-200102000-00019

Source DB:  PubMed          Journal:  Clin Orthop Relat Res        ISSN: 0009-921X            Impact factor:   4.176


  5 in total

1.  Is routine mid-term total hip arthroplasty surveillance beneficial?

Authors:  James A Keeney; Bradley S Ellison; William J Maloney; John C Clohisy
Journal:  Clin Orthop Relat Res       Date:  2012-06-06       Impact factor: 4.176

2.  Metallosis: CT findings in a total hip arthroplasty.

Authors:  Jared R Kirkham; Jonelle M Petscavage; Michael L Richardson
Journal:  Radiol Case Rep       Date:  2015-11-06

3.  An unusual presentation of catastrophic failure of hip arthroplasty with a thigh mass.

Authors:  Florian Dibra; Hari Parvataneni
Journal:  Arthroplast Today       Date:  2016-04-08

4.  UK poSt Arthroplasty Follow-up rEcommendations (UK SAFE): what does analysis of linked, routinely collected national datasets tell us about mid-late term revision risk after knee replacement?

Authors:  Lindsay K Smith; Cesar Garriga; Sarah R Kingsbury; Rafael Pinedo-Villanueva; Antonella Delmestri; Nigel K Arden; Martin Stone; Philip G Conaghan; Andrew Judge
Journal:  BMJ Open       Date:  2022-03-09       Impact factor: 2.692

5.  Prosthetic femoral head erosion through an acetabular component treated with revision and implant preservation.

Authors:  Mark Daniel Kohn; Navin Fernando
Journal:  Arthroplast Today       Date:  2020-01-14
  5 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.