Inferior vena cava loop of the implantable cardioverter defibrillator endocardial lead: a possible solution of the growth problem in pediatric implantation.

M Gasparini1, M Mantica, P Galimberti, F Coltorti, C Ceriotti, S G Priori.   

Abstract

The ICD is an important treatment option in adults and children with life-threatening tachyarrhythmias. The possibility of lead displacement caused by growth and the lack of dedicated leads and devices poses special problems in pediatric ICD implantation. We describe our experience in three children in whom we left a redundant lead loop within the inferior vena cava (IVC) is allow for further growth. Since February 1998, three children underwent ICD implantation at our institution. A lead (screw-in) was advanced into the right ventricular apex, and a loop was created in the IVC by progressively withdrawing the stylet and pushing in the lead. Satisfactory sensing and pacing threshold values were obtained and a successful single 16-J defibrillation test was performed. No complications were encountered. After a mean follow-up of 16 months, with a mean increase in body weight and height of 4.1 +/- 0.5 Kg and 6.3 +/- 0.4 cm, respectively, chest X ray showed some release of additional lead length, in the absence of dislodgments, while significant changes in pacing/sensing parameters were not found. In conclusion, the creation of a loop within the IVC allows the lead to adjust for growth in children receiving an ICD. This approach is feasible and safe.

Entities:  

Mesh:

Year:  2000        PMID: 11202255     DOI: 10.1111/j.1540-8159.2000.tb00784.x

Source DB:  PubMed          Journal:  Pacing Clin Electrophysiol        ISSN: 0147-8389            Impact factor:   1.976


  3 in total

1.  Special problems of pacing in children.

Authors:  Herwig Antretter; Joshua Colvin; Ulli Schweigmann; Herbert Hangler; Daniel Hofer; Karin Dunst; Josef Margreiter; Guenther Laufer
Journal:  Indian Pacing Electrophysiol J       Date:  2003-01-01

Review 2.  Pacing device therapy in infants and children: a review.

Authors:  Daiji Takeuchi; Yasuko Tomizawa
Journal:  J Artif Organs       Date:  2012-10-27       Impact factor: 1.731

3.  Case series of late lead dislodgement of Medtronic SelectSecure 3830 pacing leads in growing paediatric patients.

Authors:  Li Yen Ng; Sarah Gallagher; Kevin P Walsh
Journal:  Eur Heart J Case Rep       Date:  2021-01-12