N V Grassineau1, D P Mattey, D Lowry.
Abstract
A continuous flow method (CF-IRMS) for the rapid determination of the sulfur isotope composition of sulfide and sulfate minerals has significant advantages over the classic extraction method in terms of the reduced sample quantity and a rapid analytical cycle of less than 8 min/ analysis. For optimum performance, the technique is sensitive to a number of operating parameters, including sample weight and the O2 saturation of the Cu-reduction reactor. Raw data are corrected using a calibration based on five international and internal standards ranging from -17.3 to +20.3 per thousand, which requires monitoring in order to correct the effect of changing delta18O of the sample gas on the measured mass 66 values. Measured sulfur contents are within 1-1.5% of expected values and the reproducibility of delta34S values is +/-0.1 per thousand (1sigma). The technique has been used successfully for more than 1000 analyses of geological samples with a wide range of delta34S from -20 to +20 per thousand.Entities:
Year: 2001 PMID: 11199969 DOI: 10.1021/ac000550f
Source DB: PubMed Journal: Anal Chem ISSN: 0003-2700 Impact factor: 6.986