J T Wu1.
Abstract
A new parallel liquid chromatography/tandem mass spectrometry (LC/MS/MS) system has been developed, in which the mass detector was shared between two staggered parallel chromatographic runs. Since the chromatography for biofluids assay generally requires good analyte retention and thus tends to leave large blank chromatographic windows, this parallel system allowed the efficient use of the mass detector during these blank windows, resulting in significantly improved sample throughput. Also, in order to remove the bottleneck in sample extraction for this parallel separation system, a high-flow extraction device was used to perform on-line extraction. This allowed for the direct injection of biofluids onto the system. The performance and capability of this system was evaluated in tests that contained a single analyte (oxazepam) and multiple analytes (12-in-1). The results indicated that the data generated from this system were comparable to those obtained on a conventional single-column system. An application of the system for high-throughput pharmacokinetic screening of drug candidates was also demonstrated. Copyright 2001 John Wiley & Sons, Ltd.Entities:
Mesh:
Substances:
Year: 2001 PMID: 11180533 DOI: 10.1002/1097-0231(20010130)15:2<73::AID-RCM193>3.0.CO;2-Q
Source DB: PubMed Journal: Rapid Commun Mass Spectrom ISSN: 0951-4198 Impact factor: 2.419