Literature DB >> 11174285

Pitfalls in endoscope reprocessing: brushing of air and water channels is mandatory for high-level disinfection.

Y Ishino1, K Ido, H Koiwai, K Sugano.   

Abstract

BACKGROUND: Endoscopic transmission of pathogens has been reported. Guidelines have been formulated concerning the risk of infection via contaminated suction and accessory channels. Contamination of the other 2 channels for air and water has not been demonstrated. These channels were examined to clarify whether they require cleaning.
METHODS: Endoscopes used for examinations were divided into 2 groups. Group A endoscopes (n = 20) were brushed along the air and water channels. Group B endoscopes (n = 22) were not. After machine reprocessing, specimens were obtained for bacterial culture. The residual protein was measured in the 2 channels by using amido black 10B dye, and results were compared between the 2 groups.
RESULTS: With regard to the air channel, there were no contaminated endoscopes detected in either group. For the water channel, 1 endoscope in group B was positive whereas there were none positive in group A. With regard to quantification of residual protein, brushing diminished the level in both the air and the water channels.
CONCLUSION: The air and water channels can become contaminated. Brushing every channel is mandatory for high-level disinfection. A redesign of the fundamental structure of endoscopes is proposed.

Entities:  

Mesh:

Substances:

Year:  2001        PMID: 11174285     DOI: 10.1067/mge.2001.112195

Source DB:  PubMed          Journal:  Gastrointest Endosc        ISSN: 0016-5107            Impact factor:   9.427


  3 in total

Review 1.  High-level disinfection of gastrointestinal endoscope reprocessing.

Authors:  King-Wah Chiu; Lung-Sheng Lu; Shue-Shian Chiou
Journal:  World J Exp Med       Date:  2015-02-20

Review 2.  Is peracetic acid suitable for the cleaning step of reprocessing flexible endoscopes?

Authors:  Günter Kampf; Patricia M Fliss; Heike Martiny
Journal:  World J Gastrointest Endosc       Date:  2014-09-16

3.  Steps of reprocessing and equipments.

Authors:  Yong Kook Lee; Jeong Bae Park
Journal:  Clin Endosc       Date:  2013-05-31
  3 in total