Literature DB >> 11167165

Point-of-care (POC) testing of lactate in the intensive care patient. Accuracy, reliability, and costs of different measurement systems.

J Boldt1, B Kumle, S Suttner, G Haisch.   

Abstract

BACKGROUND: Point-of-care (POC) monitoring may facilitate diagnostic and therapeutic interventions in the critically ill. In a prospective study reliability and costs of two POC lactate testing devices were compared to the central laboratory lactate monitoring method.
METHODS: From arterial whole blood and capillary blood samples of 40 intensive care unit (ICU) patients lactate concentrations were measured using either a battery-powered hand-held lactate analyser (Accusport) or a bench-top blood gas analyser (Chiron 865 series) and compared with lactate measured by the hospital's central laboratory ('reference' method). Measurements were carried out on the day of inclusion in the study ('baseline'), 8 h later, and in the morning of the 1st, 2nd, and 3rd days thereafter. Costs for blood lactate measurements were analysed for each method.
RESULTS: A total of 197 lactate measurements were done with each system. Lactate measured from arterial blood using the Accusport and the central laboratory showed excellent agreement (bias: -0.15 mmol/l; relative error: -2.74%). Lactate measured by the blood gas analyser had also excellent agreement with the reference method (bias: 0.09 mmol/l; relative error: 2.24%) and the hand-held device (bias: -0.24 mmol/l; relative error: -5.06%). Time from blood sampling to availability of data was significantly longer for the central laboratory (85+/-35 min [range: 45-168 min]) than for both POC systems (range: 1-10 min). Total cost for measuring blood lactate was lowest using the Accusport device.
CONCLUSION: The trend towards miniaturizing laboratory monitoring devices will continue, resulting in greater flexibility of management of the critically ill. Blood lactate can rapidly, easily, accurately, and economically be measured at the bedside with the hand-held monitor apparatus Accusport and the bench-top blood gas analyser.

Entities:  

Mesh:

Substances:

Year:  2001        PMID: 11167165     DOI: 10.1034/j.1399-6576.2001.450210.x

Source DB:  PubMed          Journal:  Acta Anaesthesiol Scand        ISSN: 0001-5172            Impact factor:   2.105


  14 in total

1.  Temperature and time stability of whole blood lactate: implications for feasibility of pre-hospital measurement.

Authors:  Christopher W Seymour; David Carlbom; Colin R Cooke; Timothy R Watkins; Eileen M Bulger; Thomas D Rea; Geoffrey S Baird
Journal:  BMC Res Notes       Date:  2011-05-30

2.  Hemodynamic monitoring in shock and implications for management. International Consensus Conference, Paris, France, 27-28 April 2006.

Authors:  Massimo Antonelli; Mitchell Levy; Peter J D Andrews; Jean Chastre; Leonard D Hudson; Constantine Manthous; G Umberto Meduri; Rui P Moreno; Christian Putensen; Thomas Stewart; Antoni Torres
Journal:  Intensive Care Med       Date:  2007-04       Impact factor: 17.440

3.  Portable lactate analyzer for measuring lactate in cerebrospinal fluid (CSF) and plasma - method-comparison evaluations.

Authors:  Sérgio Monteiro de Almeida; Jennifer Marquie-Beck; Archana Bhatt; Scott Letendre; Allen McCutchan; Ron Ellis
Journal:  Arq Neuropsiquiatr       Date:  2014-07       Impact factor: 1.420

4.  Physiological assessment of the polytrauma patient: initial and secondary surgeries.

Authors:  N Enninghorst; R Peralta; O Yoshino; R Pfeifer; H C Pape; B M Hardy; D C Dewar; Z J Balogh
Journal:  Eur J Trauma Emerg Surg       Date:  2011-11-03       Impact factor: 3.693

5.  Validation of A Point-of-Care Lactate Device For Screening At-Risk Adults Receiving Combination Antiretroviral Therapy In Botswana.

Authors:  Sikhulile Moyo; Hermann Bussmann; Phibeon Mangwendeza; Priti Dusara; Tendani Gaolathe; Madisa Mine; Rosemary Musonda; Erik van Widenfelt; Vladimir Novitsky; Joseph Makhema; Richard G Marlink; Max Essex; C William Wester
Journal:  J Antivir Antiretrovir       Date:  2011-09-20

Review 6.  Blood lactate as a predictor for in-hospital mortality in patients admitted acutely to hospital: a systematic review.

Authors:  Ole Kruse; Niels Grunnet; Charlotte Barfod
Journal:  Scand J Trauma Resusc Emerg Med       Date:  2011-12-28       Impact factor: 2.953

7.  Theoretical predictions of lactate and hydrogen ion distributions in tumours.

Authors:  Maymona Al-Husari; Steven D Webb
Journal:  PLoS One       Date:  2013-08-21       Impact factor: 3.240

8.  Point of care blood gases with electrolytes and lactates in adult emergencies.

Authors:  Dheeraj Kapoor; Meghana Srivastava; Pritam Singh
Journal:  Int J Crit Illn Inj Sci       Date:  2014-07

9.  Occurrence and adverse effect on outcome of hyperlactatemia in the critically ill.

Authors:  Houman Khosravani; Reza Shahpori; H Thomas Stelfox; Andrew W Kirkpatrick; Kevin B Laupland
Journal:  Crit Care       Date:  2009-06-12       Impact factor: 9.097

10.  Accuracy of Handheld Point-of-Care Fingertip Lactate Measurement in the Emergency Department.

Authors:  David F Gaieski; Byron C Drumheller; Munish Goyal; Barry D Fuchs; Frances S Shofer; Kara Zogby
Journal:  West J Emerg Med       Date:  2013-02