Computer-aided surgical planning for implantation of hearing aids based on CT data in a VR environment.

F Dammann1, A Bode, E Schwaderer, M Schaich, M Heuschmid, M M Maassen.   

Abstract

A study was undertaken to assess the feasibility of a preoperative fitting test for an implantable hearing aid in a virtual reality (VR) environment. High-resolution spiral computed tomography (CT) of the mastoid bone was performed, and the results of a mastoidectomy were simulated with manual segmentation on a standard medical workstation. CT was also performed on a temporal bone specimen obtained at real mastoidectomy, and the bone margins were segmented automatically with threshold-based techniques. A triangulated surface representation of the bone structures including the mastoid cavity was generated. These data as well as the computer-aided design (CAD) files of the medical devices were transferred into a VR environment. The CAD components of the hearing aid were manipulated to simulate the surgical implantation procedure. Merging CAD data of an implantable hearing aid with CT data of the temporal bone in a VR environment was shown to be a feasible method of providing three-dimensional information for the presurgical determination of fit and mountability. Advances in hardware and software are expected to improve the usability of this method. Although clinical studies are needed, these results may serve as an impetus for exploring the use of low-cost, widely available VR computer equipment in a potentially broad field of clinical applications.

Entities:  

Mesh:

Year:  2001        PMID: 11158653     DOI: 10.1148/radiographics.21.1.g01ja21183

Source DB:  PubMed          Journal:  Radiographics        ISSN: 0271-5333            Impact factor:   5.333


  4 in total

1.  [Laser-based quality assurance for robot-assisted milling at the base of the skull].

Authors:  M M Maassen; D Malthan; J Stallkamp; A Schäfer; F Dammann; E Schwaderer; H P Zenner
Journal:  HNO       Date:  2006-02       Impact factor: 1.284

2.  Measurement error of 3D cranial landmarks of an ontogenetic sample using Computed Tomography.

Authors:  Jimena Barbeito-Andrés; Marisol Anzelmo; Fernando Ventrice; Marina L Sardi
Journal:  J Oral Biol Craniofac Res       Date:  2012-06-18

3.  Comparative use of the computer-aided angiography and rapid prototyping technology versus conventional imaging in the management of the Tile C pelvic fractures.

Authors:  Baofeng Li; Bei Chen; Ying Zhang; Xinyu Wang; Fei Wang; Hong Xia; Qingshui Yin
Journal:  Int Orthop       Date:  2015-05-21       Impact factor: 3.075

4.  Mastoid cavity dimensions and shape: method of measurement and virtual fitting of implantable devices.

Authors:  Ophir Handzel; Haobing Wang; Jason Fiering; Jeffrey T Borenstein; Mark J Mescher; Erin E Leary Swan; Brian A Murphy; Zhiqiang Chen; Marcello Peppi; William F Sewell; Sharon G Kujawa; Michael J McKenna
Journal:  Audiol Neurootol       Date:  2009-04-15       Impact factor: 1.854