| Literature DB >> 11136753 |
A van Griethuysen1, M Bes, J Etienne, R Zbinden, J Kluytmans.
Abstract
A newly marketed rapid agglutination kit for the identification of Staphylococcus aureus, Slidex Staph Plus (bioMérieux), was compared to Staphaurex Plus (Murex Diagnostics) and Pastorex Staph-Plus (Sanofi Diagnostics Pasteur). The study took place in three clinical microbiology laboratories in three different European countries. A total of 892 staphylococcal isolates, including 278 methicillin-sensitive S. aureus (MSSA) isolates, 171 methicillin-resistant S. aureus (MRSA) isolates, and 443 coagulase-negative staphylococcal isolates, were analyzed. The sensitivities (MSSA/MRSA) and specificities, respectively, were 98. 2% (98.9%/97.1%) and 98.9% for Slidex Staph Plus, 98.2% (98.2%/98. 2%) and 96.2% for Staphaurex Plus, and 98.7% (98.6%/98.8%) and 95.7% for Pastorex Staph Plus. The specificity of the Slidex Staph Plus kit was statistically significantly higher than the specificities of Staphaurex Plus and Pastorex Staph-Plus. The Slidex Staph Plus is a very reliable test for the identification of S. aureus.Entities:
Mesh:
Substances:
Year: 2001 PMID: 11136753 PMCID: PMC87684 DOI: 10.1128/JCM.39.1.86-89.2001
Source DB: PubMed Journal: J Clin Microbiol ISSN: 0095-1137 Impact factor: 5.948