Simple anterior pelvic external fixation.

M C Tucker1, S E Nork, P T Simonian, M L Routt.   

Abstract

BACKGROUND: Unstable pelvic ring disruptions are often associated with significant morbidity and mortality, especially in patients with multiple injuries. Early pelvic fixation provides stability and should diminish ongoing hemorrhage. A simple anterior single-pin pelvic external fixator can be applied rapidly and accurately to stabilize pelvic ring injuries as a part of the initial patient resuscitation of such patients. Simple anterior pelvic external fixation (SAPEF) frames can be used as either temporary, definitive, or supplementary fixation depending on the pelvic injury pattern.
METHODS: Over a 32-month period, 41 patients with unstable pelvic ring disruptions were stabilized using a simple anterior pelvic external fixator. Eight patients had open pelvic ring injuries and 13 others had genitourinary system disruptions. Fluoroscopic imaging was used to insert all of the fixation pins into the iliac crest between the iliac cortical tables to a depth of at least 5 cm. Each patient had closed manipulative reduction of the pelvic ring using external methods before SAPEF application.
RESULTS: One patient died less than 24 hours after injury because of torrential hemorrhage. Clinical evaluations and serial radiographs, including postoperative computed tomographic scans, were available for the other 40 patients postoperatively. Seventy-five of the 80 (94%) pins were completely contained between the iliac cortical tables, according to the computed tomographic scans. The initial pelvic closed reductions were maintained until the fixators were removed in 37 of 40 patients (93%). Only one deep pin track infection developed, mandating early frame removal and intravenous antibiotic therapy.
CONCLUSION: Simple anterior pelvic external fixation can be applied rapidly using fluoroscopic guidance to direct accurate pin insertion and closed manipulative reduction of the pelvis. Depending on the specific pelvic ring injury pattern and clinical scenario, SAPEF can serve as a resuscitative temporary fixation device, as definitive pelvic treatment, or as a supplement for pelvic internal fixation implants.

Entities:  

Mesh:

Year:  2000        PMID: 11130512     DOI: 10.1097/00005373-200012000-00002

Source DB:  PubMed          Journal:  J Trauma        ISSN: 0022-5282


  19 in total

1.  [Medial migration of lag screw in gamma nailing. Previously unobserved complication].

Authors:  H Thies; C von Reinhardstoettner
Journal:  Unfallchirurg       Date:  2004-08       Impact factor: 1.000

2.  Is application of an internal anterior pelvic fixator anatomically feasible?

Authors:  David J Merriman; William M Ricci; Christopher M McAndrew; Michael J Gardner
Journal:  Clin Orthop Relat Res       Date:  2012-08       Impact factor: 4.176

Review 3.  Internal versus external fixation of the anterior component in unstable fractures of the pelvic ring: pooled results from a systematic review.

Authors:  B Wardle; G D Eslick; P Sunner
Journal:  Eur J Trauma Emerg Surg       Date:  2015-08-12       Impact factor: 3.693

4.  Septic arthritis of the hip associated with supra-acetabular external fixation of unstable pelvic ring: a case report.

Authors:  Kyle T Judd; Todd O McKinley
Journal:  Iowa Orthop J       Date:  2009

5.  Midterm radiologic and functional outcomes of minimally-invasive fixation of unstable pelvic fractures using anterior internal fixator(INFIX) and percutaneous iliosacral screws.

Authors:  Ajoy Prasad Shetty; Aju Bosco; Ramesh Perumal; Jeyaramaraju Dheenadhayalan; Shanmuganathan Rajasekaran
Journal:  J Clin Orthop Trauma       Date:  2017-06-10

6.  13-Year experience in external fixation of the pelvis: complications, reduction and removal.

Authors:  P M Mitchell; C M Corrigan; N A Patel; A J Silverberg; S E Greenberg; R V Thakore; W T Obremskey; J M Ehrenfeld; J M Evans; M K Sethi
Journal:  Eur J Trauma Emerg Surg       Date:  2015-02-24       Impact factor: 3.693

7.  Complications of anterior subcutaneous internal fixation for unstable pelvis fractures: a multicenter study.

Authors:  Rahul Vaidya; Erik N Kubiak; Patrick F Bergin; Derek G Dombroski; Ren J Critchlow; Anil Sethi; Adam J Starr
Journal:  Clin Orthop Relat Res       Date:  2012-08       Impact factor: 4.176

8.  Is fixation failure after plate fixation of the symphysis pubis clinically important?

Authors:  Stephen A C Morris; Jeremy Loveridge; David K A Smart; Anthony J Ward; Tim J S Chesser
Journal:  Clin Orthop Relat Res       Date:  2012-08       Impact factor: 4.176

9.  A Guide to Improving the Care of Patients with Fragility Fractures, Edition 2.

Authors:  Simon C Mears; Stephen L Kates
Journal:  Geriatr Orthop Surg Rehabil       Date:  2015-06

10.  Definitive use of external fixation for pelvic ring injuries (open book/APC2) in pregnancy.

Authors:  V Stohlner; J R Gill; C G Murphy; A D Carrothers
Journal:  BMJ Case Rep       Date:  2015-12-16
View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.