Cylindrical thermal coagulation necrosis using an interstitial applicator with a plane ultrasonic transducer: in vitro and in vivo experiments versus computer simulations.

C Lafon1, F Prat, J Y Chapelon, F Gorry, J Margonari, Y Theillère, D Cathignol.   

Abstract

Extracorporeal HIFU techniques still cannot be used to treat tumours of the digestive tract, therefore an interstitial applicator has been developed to fill this gap. The object of the study was to validate the use of a plane ultrasonic transducer in an interstitial applicator to obtain large sector based or cylindrical coagulation necrosis. Two very different shot sequences were performed in vitro and in vivo and compared with numerical calculations. Each sequence consisted of 20 shots. After each shot the applicator was rotated through an angle of 18 degrees. Each shot in sequence 1 lasted 20 s, with a 2-min interval between shots which can be considered as independent. The second sequence involved coupling the shots so that each benefits from the heat deposited by the preceding ones. The first shot lasted 20 s to establish the lesion, then the duration of the subsequent 19 shots was 10 s to take into account the temperature rise due to preceding shots. In both cases, it was shown that cylindrical necrosis resulted in vivo and in vitro: 20 mm diameter and 8mm in height. The dimensions and the shapes of the necrosed volumes agreed with numerical predictions: the necrosed area induced by sequence 1 had a serrated border, whereas that of the second sequence was much more uniform. It was also shown that, for the two sequences, less than 20 s were necessary to coagulate the tissues in each direction. The results with sequence 2 showed that coupling the shots could be used to reduce treatment time without modifying the necrosed volume.

Entities:  

Mesh:

Year:  2000        PMID: 11129262     DOI: 10.1080/02656730050199359

Source DB:  PubMed          Journal:  Int J Hyperthermia        ISSN: 0265-6736            Impact factor:   3.914


  6 in total

1.  A 2D fast near-field method for calculating near-field pressures generated by apodized rectangular pistons.

Authors:  Duo Chen; Robert J McGough
Journal:  J Acoust Soc Am       Date:  2008-09       Impact factor: 1.840

2.  FFT-based computation of the bioheat transfer equation for the HCC ultrasound surgery therapy modeling.

Authors:  Jean-Louis Dillenseger; Simon Esneault; Carole Garnier
Journal:  Conf Proc IEEE Eng Med Biol Soc       Date:  2008

3.  Theoretical comparison of two interstitial ultrasound applicators designed to induce cylindrical zones of tissue ablation.

Authors:  C Lafon; F Chavrier; F Prat; J Y Chapelon; D Cathignol
Journal:  Med Biol Eng Comput       Date:  1999-05       Impact factor: 2.602

4.  Considerations for theoretical modelling of thermal ablation with catheter-based ultrasonic sources: implications for treatment planning, monitoring and control.

Authors:  Punit Prakash; Chris J Diederich
Journal:  Int J Hyperthermia       Date:  2012       Impact factor: 3.914

5.  3-D modeling of the thermal coagulation necrosis induced by an interstitial ultrasonic transducer.

Authors:  Carole Garnier; Cyril Lafon; Jean-Louis Dillenseger
Journal:  IEEE Trans Biomed Eng       Date:  2008-02       Impact factor: 4.538

Review 6.  Modelling of endoluminal and interstitial ultrasound hyperthermia and thermal ablation: applications for device design, feedback control and treatment planning.

Authors:  Punit Prakash; Vasant A Salgaonkar; Chris J Diederich
Journal:  Int J Hyperthermia       Date:  2013-06       Impact factor: 3.914

  6 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.