Literature DB >> 11108634

Design and fabrication of a three-axis edge ROU head and neck gradient coil.

B A Chronik1, A Alejski, B K Rutt.   

Abstract

The design, fabrication, and testing of a complete three-axis gradient coil capable of imaging the human neck is described. The analytic method of constrained current minimum inductance (CCMI) was used to position the uniform region of the gradient coil adjacent to and extending beyond the physical edge of the coil. The average gradient efficiency of the three balanced axes is 0.37 mT/m/A and the average inductance is 827 microH. With maximum amplifier current of 200A and receive signal sweep width of +/-125 kHz, the average minimum FOV using this gradient set is 7.9 cm. The completed coil has an inner diameter of 32 cm, an outer diameter of 42 cm, and a length (including cabling connections) of 80 cm. The entire coil was built in-house. The structure is actively water cooled. Heating measurements were made to characterize the thermal response of the coil under various operating conditions and it was determined that a continuous current of 100A could be passed through all three axes simultaneously without increasing the internal coil temperature by more than 23 degrees C. Eddy current measurements were made for all axes. With digital compensation, the gradient eddy current components could be adequately compensated. A large B(o) eddy current field is produced by the Gz axis that could be corrected through the use of an auxiliary B(o) compensation coil. Preliminary imaging results are shown in both phantoms and human subjects.

Entities:  

Mesh:

Year:  2000        PMID: 11108634     DOI: 10.1002/1522-2594(200012)44:6<955::aid-mrm18>3.0.co;2-1

Source DB:  PubMed          Journal:  Magn Reson Med        ISSN: 0740-3194            Impact factor:   4.668


  19 in total

1.  New head gradient coil design and construction techniques.

Authors:  William B Handler; Chad T Harris; Timothy J Scholl; Dennis L Parker; K Craig Goodrich; Brian Dalrymple; Frank Van Sass; Blaine A Chronik
Journal:  J Magn Reson Imaging       Date:  2013-10-09       Impact factor: 4.813

2.  High slew-rate head-only gradient for improving distortion in echo planar imaging: Preliminary experience.

Authors:  Ek T Tan; Seung-Kyun Lee; Paul T Weavers; Dominic Graziani; Joseph E Piel; Yunhong Shu; John Huston; Matt A Bernstein; Thomas K F Foo
Journal:  J Magn Reson Imaging       Date:  2016-02-26       Impact factor: 4.813

3.  On the accurate analysis of vibroacoustics in head insert gradient coils.

Authors:  Simone A Winkler; Andrew Alejski; Trevor Wade; Charles A McKenzie; Brian K Rutt
Journal:  Magn Reson Med       Date:  2016-11-17       Impact factor: 4.668

4.  Lightweight, compact, and high-performance 3T MR system for imaging the brain and extremities.

Authors:  Thomas K F Foo; Evangelos Laskaris; Mark Vermilyea; Minfeng Xu; Paul Thompson; Gene Conte; Christopher Van Epps; Christopher Immer; Seung-Kyun Lee; Ek T Tan; Dominic Graziani; Jean-Baptise Mathieu; Christopher J Hardy; John F Schenck; Eric Fiveland; Wolfgang Stautner; Justin Ricci; Joseph Piel; Keith Park; Yihe Hua; Ye Bai; Alex Kagan; David Stanley; Paul T Weavers; Erin Gray; Yunhong Shu; Matthew A Frick; Norbert G Campeau; Joshua Trzasko; John Huston; Matt A Bernstein
Journal:  Magn Reson Med       Date:  2018-03-13       Impact factor: 4.668

5.  Peripheral nerve stimulation characteristics of an asymmetric head-only gradient coil compatible with a high-channel-count receiver array.

Authors:  Seung-Kyun Lee; Jean-Baptiste Mathieu; Dominic Graziani; Joseph Piel; Eric Budesheim; Eric Fiveland; Christopher J Hardy; Ek Tsoon Tan; Bruce Amm; Thomas K-F Foo; Matt A Bernstein; John Huston; Yunhong Shu; John F Schenck
Journal:  Magn Reson Med       Date:  2015-12-02       Impact factor: 4.668

6.  Hyperpolarized (13)C-lactate to (13)C-bicarbonate ratio as a biomarker for monitoring the acute response of anti-vascular endothelial growth factor (anti-VEGF) treatment.

Authors:  Jae Mo Park; Daniel M Spielman; Sonal Josan; Taichang Jang; Milton Merchant; Ralph E Hurd; Dirk Mayer; Lawrence D Recht
Journal:  NMR Biomed       Date:  2016-03-14       Impact factor: 4.044

7.  Gradient pre-emphasis to counteract first-order concomitant fields on asymmetric MRI gradient systems.

Authors:  Shengzhen Tao; Paul T Weavers; Joshua D Trzasko; Yunhong Shu; John Huston; Seung-Kyun Lee; Louis M Frigo; Matt A Bernstein
Journal:  Magn Reson Med       Date:  2016-07-04       Impact factor: 4.668

8.  Peripheral nerve stimulation limits of a high amplitude and slew rate magnetic field gradient coil for neuroimaging.

Authors:  Ek T Tan; Yihe Hua; Eric W Fiveland; Mark E Vermilyea; Joseph E Piel; Keith J Park; Vincent B Ho; Thomas K F Foo
Journal:  Magn Reson Med       Date:  2019-08-06       Impact factor: 4.668

9.  Reduced acoustic noise in diffusion tensor imaging on a compact MRI system.

Authors:  Ek T Tan; Christopher J Hardy; Yunhong Shu; Myung-Ho In; Arnaud Guidon; John Huston; Matt A Bernstein; Thomas K F Foo
Journal:  Magn Reson Med       Date:  2017-10-02       Impact factor: 4.668

10.  Clinical field-strength MRI of amyloid plaques induced by low-level cholesterol feeding in rabbits.

Authors:  John A Ronald; Yuanxin Chen; Lisa Bernas; Hagen H Kitzler; Kem A Rogers; Robert A Hegele; Brian K Rutt
Journal:  Brain       Date:  2009-03-17       Impact factor: 13.501

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.