Literature DB >> 11090728

Clinical impact of hypnotic-titration guidelines based on EEG bispectral index (BIS) monitoring during routine anesthetic care.

J W Johansen1, P S Sebel, J C Sigl.   

Abstract

STUDY
OBJECTIVES: To examine the impact on perioperative care of routine Bispectral Index (BIS) monitoring during general anesthesia throughout an entire operating room (OR) suite.
DESIGN: Open, observational trial with retrospective analysis of guideline performance. Data were analyzed from 1,552 adult patients receiving general anesthesia with surgical times of at least 1 hour and who were extubated by postanesthesia care unit (PACU) discharge. Staff were trained using a simple decision matrix, which integrated BIS titration goals with anesthetic management. Unmonitored patients were compared to either BIS-monitored patients or to performance subgroups based on BIS measurements recorded during anesthetic maintenance ("deep", BIS < 50; "target", 50-65; "light", >65).
SETTING: Large, urban academic/trauma center.
MEASUREMENTS AND MAIN RESULTS: Demographic profiles of all groups and subgroups were similar. Anesthetic emergence, recovery times, and volatile drug use were significantly shortened or reduced only when BIS values were maintained between 50 and 65. Extubation time from end of surgery decreased by 2.1 minutes from 5.7+/-7 (37%); OR exit time decreased by 2.2 minutes from 9.3+/-6 (24%); eligibility for phase 1 PACU discharge decreased by 4 minutes from 22+/-42 (23%); and actual PACU discharge decreased by 15 minutes from 130+/-78 (7%). PACU extubation frequency decreased from 6.9% to 2.6%. Modest decreases in total intraoperative drug use were noted with an increase in PACU analgesic administration.
CONCLUSIONS: Routine application of BIS monitoring throughout an OR suite impacted clinical outcome only if guideline targets were met. BIS values within the last 30 minutes of surgery were not predictive of emergence or recovery. Hypnotic maintenance at BIS < 50 did not confer any clinical advantage over unmonitored cases. Anesthetic maintenance at BIS values between 50 and 65 was associated with shortened emergence and recovery from general anesthesia.

Entities:  

Mesh:

Year:  2000        PMID: 11090728     DOI: 10.1016/s0952-8180(00)00187-2

Source DB:  PubMed          Journal:  J Clin Anesth        ISSN: 0952-8180            Impact factor:   9.452


  13 in total

1.  Comparison of the EEG-based SNAP index and the Bispectral (BIS) index during sevoflurane-nitrous oxide anaesthesia.

Authors:  P Ruiz-Gimeno; M Soro; A Pérez-Solaz; M Carrau; F J Belda; J L Jover; G Aguilar
Journal:  J Clin Monit Comput       Date:  2006-01-25       Impact factor: 2.502

2.  Usefulness of permutation entropy as an anesthetic depth indicator in children.

Authors:  Pil-Jong Kim; Hong-Gee Kim; Gyu-Jeong Noh; Yong-Seo Koo; Teo Jeon Shin
Journal:  J Pharmacokinet Pharmacodyn       Date:  2015-02-04       Impact factor: 2.745

3.  Multi-scale sample entropy of electroencephalography during sevoflurane anesthesia.

Authors:  Yinghua Wang; Zhenhu Liang; Logan J Voss; Jamie W Sleigh; Xiaoli Li
Journal:  J Clin Monit Comput       Date:  2014-01-11       Impact factor: 2.502

4.  [Influence of gender on propofol consumption and recovery times].

Authors:  W Wilhelm; H Buchinger; A Biedler; S Altmann; R Larsen; S Kreuer
Journal:  Anaesthesist       Date:  2005-06       Impact factor: 1.041

Review 5.  Bispectral index for improving anaesthetic delivery and postoperative recovery.

Authors:  Yodying Punjasawadwong; Aram Phongchiewboon; Nutchanart Bunchungmongkol
Journal:  Cochrane Database Syst Rev       Date:  2014-06-17

6.  Bispectral index monitor: an evidence-based analysis.

Authors: 
Journal:  Ont Health Technol Assess Ser       Date:  2004-06-01

7.  Correlation beween AAI-index and the BIS-index during propofol hypnosis: a clinical study.

Authors:  Russell E Anderson; Gunilla Barr; Jan G Jakobsson
Journal:  J Clin Monit Comput       Date:  2002-08       Impact factor: 2.502

8.  Validation of the bispectral index as an indicator of anesthetic depth in Thoroughbred horses anesthetized with sevoflurane.

Authors:  Hirotaka Tokushige; Masashi Kakizaki; Hirotaka Ode; Atsushi Okano; Jun Okada; Taisuke Kuroda; Ai Wakuno; Minoru Ohta
Journal:  J Equine Sci       Date:  2016-12-15

9.  Effects of bispectral index monitoring on isoflurane consumption and recovery profiles for anesthesia in an elderly asian population.

Authors:  Faraz Shafiq; Hamid Iqil Naqvi; Aliya Ahmed
Journal:  J Anaesthesiol Clin Pharmacol       Date:  2012-07

10.  Comparison of recovery profiles of propofol and sevoflurane anesthesia with bispectral index monitoring in percutaneous nephrolithotomy.

Authors:  Zeynep Nur Orhon; Sibel Devrim; Melek Celik; Yekbun Dogan; Asif Yildirim; Erem Kaan Basok
Journal:  Korean J Anesthesiol       Date:  2013-03-19
View more