Literature DB >> 11074515

Epidermal protection with cryogen spray cooling during high fluence pulsed dye laser irradiation: an ex vivo study.

J W Tunnell1, J S Nelson, J H Torres, B Anvari.   

Abstract

BACKGROUND AND
OBJECTIVE: Higher laser fluences than currently used in therapy (5-10 J/cm(2)) are expected to result in more effective treatment of port wine stain (PWS) birthmarks. However, higher incident fluences increase the risk of epidermal damage caused by absorption of light by melanin. Cryogen spray cooling offers an effective method to reduce epidermal injury during laser irradiation. The objective of this study was to determine whether high laser incident fluences (15-30 J/cm(2)) could be used while still protecting the epidermis in ex vivo human skin samples. STUDY DESIGN/
MATERIALS AND METHODS: Non-PWS skin from a human cadaver was irradiated with a Candela ScleroPlus Laser (lambda = 585 nm; pulse duration = 1.5 msec) by using various incident fluences (8-30 J/cm(2)) without and with cryogen spray cooling (refrigerant R-134a; spurt durations: 40-250 msec). Assessment of epidermal damage was based on histologic analysis.
RESULTS: Relatively short spurt durations (40-100 msec) protected the epidermis for laser incident fluences comparable to current therapeutic levels (8-10 J/cm(2)). However, longer spurt durations (100-250 msec) increased the fluence threshold for epidermal damage by a factor of three (up to 30 J/cm(2)) in these ex vivo samples.
CONCLUSION: Results of this ex vivo study show that epidermal protection from high laser incident fluences can be achieved by increasing the cryogen spurt duration immediately before pulsed laser exposure. Copyright 2000 Wiley-Liss, Inc.

Entities:  

Mesh:

Substances:

Year:  2000        PMID: 11074515     DOI: 10.1002/1096-9101(2000)27:4<373::aid-lsm11>3.0.co;2-a

Source DB:  PubMed          Journal:  Lasers Surg Med        ISSN: 0196-8092            Impact factor:   4.025


  2 in total

1.  Preliminary investigations on therapy thresholds for laser dosimetry, cryogen spray cooling duration, and treatment cycles for laser cartilage reshaping in the New Zealand white rabbit auricle.

Authors:  Cara A Chlebicki; Dmitry E Protsenko; Brian J Wong
Journal:  Lasers Med Sci       Date:  2013-11-08       Impact factor: 3.161

2.  Minimizing superficial thermal injury using bilateral cryogen spray cooling during laser reshaping of composite cartilage grafts.

Authors:  Cheng-Jen Chang; Sally M H Cheng; Lynn L Chiu; Brian J F Wong; Keen Ting
Journal:  Lasers Surg Med       Date:  2008-09       Impact factor: 4.025

  2 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.