Literature DB >> 11056754

Lack of agreement between tonometric and gastric juice partial carbon dioxide tension.

A Dubin1, J Badie, S Fernandez, E Estenssoro, H Canales, G Bordoli, F Pálizas.   

Abstract

STATEMENT OF
FINDINGS: Our goal was to compare measurement of tonometered saline and gastric juice partial carbon dioxide tension (PCO2). In this prospective observational study, 112 pairs of measurements were simultaneously obtained under various hemodynamic conditions, in 15 critical care patients. Linear regression analysis showed a significant correlation between the two methods of measuring PCO2 (r(2) = 0.43; P < 0.0001). However, gastric juice PCO2 was systematically higher (mean difference 51 mmHg). The 95% limits of agreement were 315 mmHg and the dispersion increased as the values of PCO2 increased. Tonometric and gastric juice PCO2 cannot be used interchangeably. Gastric juice PCO2 measurement should be interpreted with caution.

Entities:  

Mesh:

Substances:

Year:  2000        PMID: 11056754      PMCID: PMC29045          DOI: 10.1186/cc701

Source DB:  PubMed          Journal:  Crit Care        ISSN: 1364-8535            Impact factor:   9.097


Introduction

In recent years there has been growing interest in tonometric estimation of gastric pHi. More recently attention has focused on the gradient between intraluminal and arterial PCO2. pHi appears to be a useful diagnostic and prognostic tool in critically ill patients, and may also be used as a therapeutic guide [2,3]. However, intraluminal PCO2 is the parameter measured to calculate pHi, and it is assumed as equivalent to the PCO2 of the upper layers of the gastric mucosa [4]. Direct measurement of PCO2 in gastric juice might offer additional advantages over tonometry. First, tonometer costs could be saved. Second, equilibration time would no longer be necessary. Finally, preanalytic factors that account for poor tonometric reproducibility, such as inadequate volume of saline in the tonometer, errors in timing the dwell time of the sample or in the technique used to aspirate the tonometered saline sample, mixing the sample with air from the tonometer dead space, and delays in specimen analysis, might be prevented [5]. Nevertheless, to our knowledge, very few studies, either experimental or clinical, have examined PCO2 in gastric juice [1,6,7]. Moreover, no comparison with tonometric samples obtained at the same time has been done. Our goal was to compare PCO2 obtained simultaneously in gastric juice and in saline samples from tonometers. The results of the present study show that gastric juice PCO2 is systematically higher, and that for high PCO2 values this difference widens and data dispersion is even more marked. Therefore, tonometric PCO2 and gastric juice PCO2 are not interchangeable.

Patients and methods

The present study was approved by the local ethics committee and informed consent was obtained from the next of kin of each patient. We consecutively studied 15 mechanically ventilated patients from a medical/surgical intensive care unit, in whom tonometric monitoring was indicated by attending physicians. All patients were receiving 50 mg intravenous ranitidine every 8 h. Gastric tonometers were filled with saline and, after 90 min for equilibration, saline samples were collected, as has previously been described [2]. At the same time, gastric juice was anaerobically extracted from the aspiration port of the tonometer. Initial 20 ml were discarded. A blood gas analyzer (AVL 945) was used to measure PCO2 in both samples. These measurements were taken at various time points in each patient, and under various haemodynamic and oxygen transport conditions. All measurements were performed with the patient fasted. Correlation between both of measurements was examined using the Bland-Altman technique [8]. We also performed an in vitro study to quantify the precision and bias of the AVL 945. For this purpose, a stable PCO2 in saline solution was achieved by bubbling 5% carbon dioxide calibration gas. Bias (mean difference from expected PCO2) and precision (standard deviation of the bias) of PCO2 measured in saline was determined by comparison of measured PCO2 with expected PCO2. The latter was calculated from the carbon dioxide content of the calibration gas and from barometric pressure, according to gas laws. Measurements were repeated six times.

Results

We performed 112 pairs of measurements in 15 patients. Table 1 shows clinical data and the first values of arterial, tonometer and gastric juice PCO2 taken in each patient. Regression analysis demonstrated a significant correlation between both methods of measuring PCO2(r 2 =0.43; gastric juice PCO2 = -28.79+ [2.55 × tonometric PCO2]; P < 0.0001; Fig. 1). However, bias, calculated as the mean difference between gastric juice and tonometric PCO2, was 51 mmHg. The 95% limits of agreement were 315 mmHg (Fig. 2). When mean PCO2 values were less than 100 mmHg, bias and 95% limits of agreement were 19 and 102 mmHg, respectively. The scattering of differences widened as PCO2 increased (r 2 =0.71; P < 0.0001).
Table 1

Clinical characteristics and first value of arterial, tonometer and gastric juice PCO2

PCO2 (mmHg)

AgeInotropesGastric
PatientSex(years)Diagnosis(μg/kg per min)OutcomeArterialTonometricjuice
1Female53Stroke, ARDSDopamine 32Survival3048165
2Female73Intestinal obstruction, septic shockDopamine 40Death264492
3Male37Multiple traumaSurvival212841
4Male56Multiple traumaSurvival394249
5Female64Acute pancreatitis, shock, ARDSDopamine 18Death303480
6Male17Multiple traumaSurvival436060
7Female18Fat liver of pregnancySurvival304044
8Male73Necrotizing celulitis, septic shock, ARDSEpinephrine 1.2Death283331
9Male64Multiple trauma, pneumonia, ARDSDeath364157
10Male65Lung cancer postoperatively, ARDSDeath3551242
11Male65Lung cancer postoperatively, ARDSDopamine 20Death3630125
12Female22Neutropenia, septic shock, ARDSEpinephrine 0.8Death506981
13Male83Perioperative shockDopamine 25Survival232834
14Male52Ventilator-associated pneumoniaSurvival4343126
15Male56Colangitis, septic shockDopamine 36Survival384492

ARDS, acute respiratory distress syndrome.

Figure 1

Correlation between gastric juice and tonometric PCO2. We performed 112 pairs of measurements of gastric juice and tonometric PCO2 in 15 critical care patients under different haemodynamic and oxygen transport conditions. The linear regression coefficient is significant. However, the slope value indicates systematic overestimation of gastric juice PCO2 in relation to saline PCO2.

Figure 2

Bland-Altman analysis of the differences between gastric juice and tonometric PCO2. The bias calculated as the mean difference of gastric juice and tonometric PCO2 was 51 mmHg. The 95% limits of agreement were 315 mmHg. The bias and the scattering of differences widened as PCO2 increased.

The two types of measurements are clearly correlated. However, they cannot be considered as interchangeable, because gastric juice PCO2 was always higher and the 95% limits of agreement were clinically significant. In an effort to prevent bias related to multiple measurements per patient we used another approach, taking into account the initial measurement of each patient. Despite this, results continued to be similar (bias 55 mmHg, 95% limits of agreement 216 mmHg). The AVL 945 blood gas analyzer showed a negative bias of 0.97 mmHg and a precision of 2.13 mmHg. This was considered negligible, so no further correction was done to tonometric values.

Discussion

The present data show that tonometric PCO2 and gastric juice PCO2 should not be considered interchangeable,because gastric juice PCO2 was systematically higher. With high PCO2 values, this difference widened, and data dispersion became even more marked. There is no clear cause for these observations. It can be argued that tonometric PCO2 is a value that is measured over a predetermined period, whereas gastric juice PCO2 may generate data on minute-to-minute changes in mucosal metabolism. This different equilibrium time could account for data dispersion. However, the positive bias for gastric juice is harder to interpret, because such rapid changes should appear in both directions. Another potential confounding factor is the ability of blood gas analyzers to measure PCO2 in gastric juice. Measurement of PCO2 in saline is an important source of error in the estimation of pHi. The error lies both in the kind of blood gas analyzer used and in the PCO2 value itself [9,10]. Different solutions might modify PCO2 measurement. For example, bias is -66.5% when the Nova Stat Profile 7 blood gas analyzer measures concentration of 1.95% of CO2 equilibrated in normal saline. However, bias changes to +45.4% when 1.95% CO2 is equilibrated in human albumin solution 4.5% [9]. It would not be surprising if different gastric juice buffers, such as proteins, mucopolisaccharides and others, interfere with CO2 solubility and its subsequent measurement. In this way, intersubject and intrasubject variation of gastric juice composition could also account for data dispersion. Therefore, analytic issues related to the various constituents of gastric juice could have added to the observed differences. Fiddian-Green et al [1] measured PCO2 in gastric content of anaesthetised dogs. They isolated the stomach from the oesophagus and the duodenum with ligatures, and washed it through a catheter with saline. Then they instilled 250 ml saline and intermittently took samples to measure PCO2 and estimate pHi, which was compared with simultaneous direct mucosa pHi measurement performed using a microglass probe. They found a statistically significant correlation between both methods, but there was considerable data dispersion in the graph. However, in that study, as well as in others in which PCO2 was measured without a tonometer [6,7], some kind of saline lavage was used. On the other hand, we directly measured PCO2 in gastric juice, which could produce marked differences in the results. Differences between tonometric PCO2 and gastric juice PCO2 could be due to blood gas analyzers that are calibrated for blood gas analysis, and could systematically underestimate PCO2 in saline [9,10]. Nevertheless, this explanation is not supported by the present in vitro results. The performance of the AVL 945 was fairly good. It showed a minor negative bias of less than 1 mmHg and a precision of approximately 2 mmHg. Thus, it appears to be a suitable device for saline tonometry. We ascribe the high PCO2 values obtained in the present study to shock and subsequent gut hypoperfusion, as can be deduced from the clinical characteristics of our patients. Also, in the presence of hydrochloric acid, duodenal or gastric bicarbonate buffering could produce important amounts of carbon dioxide [11]. Although we cannot exclude spurious generation of carbon dioxide, measurements were performed in the fasted state, the position of the tonometer was checked by X-ray, and ranitidine was administered in each patient. If PCO2 in gastric juice and in the tonometer were interchangeable, their values would be similar, regardless of CO2 source or absolute value. However, gastric juice PCO2 and tonometric PCO2 were quite different. We believe that this is the novel point in this study. Gastric juice PCO2 has been used in some clinical studies [6,7,12]. Mohsenifar et al [12] advocated its advantages over tonometry. However, the present results suggest that the two techniques are not interchangeable. From the present data we cannot infer which should be the gold standard technique for measurement of gastric mucosa PCO2. Nevertheless, knowledge regarding pHi has evolved from gastric tonometry; studies regarding normal values, prognostic changes and its uses as a therapeutic guide were performed using tonometry. Hence, until more and clearer data are reported, PCO2 measurement in gastric juice should be considered with caution.
  12 in total

1.  Statistical methods for assessing agreement between two methods of clinical measurement.

Authors:  J M Bland; D G Altman
Journal:  Lancet       Date:  1986-02-08       Impact factor: 79.321

2.  Prognostic value of gastric intramural pH in surgical intensive care patients.

Authors:  T Gys; A Hubens; H Neels; L F Lauwers; R Peeters
Journal:  Crit Care Med       Date:  1988-12       Impact factor: 7.598

3.  Validation of tonometric measurement of gut intramural pH during endotoxemia and mesenteric occlusion in pigs.

Authors:  J B Antonsson; C C Boyle; K L Kruithoff; H L Wang; E Sacristan; H R Rothschild; M P Fink
Journal:  Am J Physiol       Date:  1990-10

4.  Measuring carbon dioxide tension in saline and alternative solutions: quantification of bias and precision in two blood gas analyzers.

Authors:  D Riddington; B Venkatesh; T Clutton-Brock; J Bion; K B Venkatesh
Journal:  Crit Care Med       Date:  1994-01       Impact factor: 7.598

5.  Back-diffusion of CO2 and its influence on the intramural pH in gastric mucosa.

Authors:  R G Fiddian-Green; G Pittenger; W M Whitehouse
Journal:  J Surg Res       Date:  1982-07       Impact factor: 2.192

6.  Mechanism for high PCO2 in gastric juice: roles of bicarbonate secretion and CO2 diffusion.

Authors:  M H Stevens; R C Thirlby; M Feldman
Journal:  Am J Physiol       Date:  1987-10

7.  Gastric intramural pH as a predictor of success or failure in weaning patients from mechanical ventilation.

Authors:  Z Mohsenifar; A Hay; J Hay; M I Lewis; S K Koerner
Journal:  Ann Intern Med       Date:  1993-10-15       Impact factor: 25.391

8.  Gastric intramucosal pH as a therapeutic index of tissue oxygenation in critically ill patients.

Authors:  G Gutierrez; F Palizas; G Doglio; N Wainsztein; A Gallesio; J Pacin; A Dubin; E Schiavi; M Jorge; J Pusajo
Journal:  Lancet       Date:  1992-01-25       Impact factor: 79.321

9.  Gastric mucosal pH as a prognostic index of mortality in critically ill patients.

Authors:  G R Doglio; J F Pusajo; M A Egurrola; G C Bonfigli; C Parra; L Vetere; M S Hernandez; S Fernandez; F Palizas; G Gutierrez
Journal:  Crit Care Med       Date:  1991-08       Impact factor: 7.598

10.  Saline PCO2 is an important source of error in the assessment of gastric intramucosal pH.

Authors:  J Takala; I Parviainen; M Siloaho; E Ruokonen; E Hämäläinen
Journal:  Crit Care Med       Date:  1994-11       Impact factor: 7.598

View more
  1 in total

Review 1.  Tonometry of partial carbon dioxide tension in gastric mucosa: use of saline, buffer solutions, gastric juice or air.

Authors:  A J Groeneveld
Journal:  Crit Care       Date:  2000-06-20       Impact factor: 9.097

  1 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.