Abstract
BACKGROUND: Measurement of cardiac output and extravascular lung water in critically ill patients using femoral artery double-indicator dilution involves femoral artery catheterization. The potential risk of vascular compromise to the limb may be exacerbated in patients receiving vasopressors. The utility of scanning laser Doppler flowmetry to measure changes in pedal perfusion following catheterization was assessed.Entities:
Year: 1999 PMID: 11056731 PMCID: PMC29021 DOI: 10.1186/cc350
Source DB: PubMed Journal: Crit Care ISSN: 1364-8535 Impact factor: 9.097
Figure 1Laser Doppler scans of the sole of the foot. The photographic image and colour coded perfusion map were made simultaneously.
Systemic arterial and calf occlusion pressures (mmHg) prior to, immediately after and 24 h following arterial catheterization
| Preinsertion | Postinsertion | 24 h | |||
| Mean systemic arterial pressure | 76 ± 11 | 80 ± 12 | 0.11 | 79 ± 11 | 0.44 |
| Catheter leg maximum occlusion pressure | 112 ± 21 | 108 ± 27 | 0.41 | 117 ± 26 | 0.65 |
| Noncatheter leg maximum occlusion pressure | 111 ± 33 | 111 ± 28 | 0.89 | 114 ± 18 | 0.82 |
Values are shown as mean± SD. P values of postcatheterization and 24 h measurements are with regard to preinsertion values.
Foot cutaneous perfusion prior to, immediately following and 24 h after arterial catheterization
| Preinsertion | Postinsertion | 24 h | |
| Insertion leg | 230.5 ± 173 | 205.7 ± 178 | 217.1 ± 146 |
| Noninsertion leg | 218.1 ± 170 | 188.8 ± 176 | 219.9 ± 171 |
| Insertion minus noninsertion legs | 12.5 ± 72 | 16.8 ± 51 | -2.9 ± 77 |
| — | 0.22 | 0.80 |
Values are shown as mean ± SD. P values of postcatheterization and 24 h measurements are with regard to preinsertion values.
Figure 2Schematic diagram of scanning laser Doppler technique. The computer directs the mirror to scan the laser beam across the predesignated area. Reflected Doppler shifted light is decoded by the signal processor which then creates a colour-coded perfusion image on the computer screen. The whole image or areas within it can be outlined and perfusion within them automatically calculated.
Figure 3Schematic drawing of scanning laser Doppler assembly on its mobile stand.