Abstract

A simple and inexpensive method was developed to obtain a coaxial view of the pedicles to assist with screw insertion. The authors evaluated the accuracy of this device to place transpedicular vertebral screws in a human adult cadaver model. A dual radiation targeting system, a laser targeting system for fluoroscopically guided procedures, was developed to provide an accurate surface entry point and angle of approach to radiographic landmarks. After fluoroscopic cross-hair target localization of the coaxial view of the pedicle, X-ray radiation is turned off and the laser beam allows the surgeon to guide the screw through the pedicle. Nine cadaver spines were removed and mounted. Three surgeons, inexperienced in the technique of pedicle screw placement, fitted instruments to 184 pedicles between L5 and T5. A total of 83 lumbar and 101 thoracic pedicles underwent screw placement. After specimen dissection, the degree and location of any screw perforation were measured by direct inspection. Three screws perforated a pedicle, for an error rate of 1.6%. Two lumbar screws (2.4% error) and one thoracic screw (1% error) perforated the pedicle. No screw was more than 1 mm outside the pedicle. Five other screws, four in the thoracic spine and one in the lumbar spine (error rate of 2.7%) were directed too far laterally and perforated the lateral vertebral body. This low rate of pedicle wall cortical perforation by inexperienced surgeons compares favorably with much higher pedicle perforation rates by experienced surgeons when no imaging was used. In conclusion, this in vitro model using a dual radiation targeting system assisted with the accurate placement of transpedicular vertebral screws with minimal radiation exposure.

Entities:  

Mesh:

Year:  2000        PMID: 11052350     DOI: 10.1097/00002517-200010000-00007

Source DB:  PubMed          Journal:  J Spinal Disord        ISSN: 0895-0385


  3 in total

1.  Image-guided pedicle screw insertion accuracy: a meta-analysis.

Authors:  Nai-Feng Tian; Hua-Zi Xu
Journal:  Int Orthop       Date:  2009-05-08       Impact factor: 3.075

2.  Reduction of radiation dose during facet joint injection using the new image guidance system SabreSource: a prospective study in 60 patients.

Authors:  Dirk Proschek; K Kafchitsas; M A Rauschmann; A A Kurth; T J Vogl; Florian Geiger
Journal:  Eur Spine J       Date:  2008-12-11       Impact factor: 3.134

3.  The substantiation of the elastic-viscoplastic model of the human spine for modeling the correction process of kyphoscoliotic deformation.

Authors:  Konstantin S Sergeev; Valery V Piven
Journal:  J Craniovertebr Junction Spine       Date:  2018 Jan-Mar
  3 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.