Anaerobic specimen transport device.

T D Wilkins, F Jimenez-Ulate.   

Abstract

A device is described and evaluated for the anaerobic transport of clinical specimens. The device limits the amount of oxygen entering with the sample to a maximum of 2%, which is rapidly removed by reacting with hydrogen in the presence of a palladium catalyst. The viability on swabs of 12 species of anaerobes, four strains of facultative anaerobes and a strain of Pseudomonas aeruginosa, was maintained during the length of the tests (24 or 48 h). The results demonstrated that this device protected even the more oxygen-sensitive clinical anaerobes from death due to oxygen exposure. This device can be used for swabs as well as for anaerobic collection and liquid and solid specimens.

Entities:  

Mesh:

Substances:

Year:  1975        PMID: 1104656      PMCID: PMC274205          DOI: 10.1128/jcm.2.5.441-447.1975

Source DB:  PubMed          Journal:  J Clin Microbiol        ISSN: 0095-1137            Impact factor:   5.948


  2 in total

1.  Use of anaerobic glove boxes for the cultivation of strictly anaerobic bacteria.

Authors:  A Aranki; R Freter
Journal:  Am J Clin Nutr       Date:  1972-12       Impact factor: 7.045

2.  Efficiency of a transport medium for the recovery of aerobic and anaerobic bacteria from applicator swabs.

Authors:  A L Barry; G D Fay; R L Sauer
Journal:  Appl Microbiol       Date:  1972-07
  2 in total
  7 in total

1.  Evaluation of Port-A-Cul transport system for protection of anaerobic bacteria.

Authors:  E Mena; F S Thompson; A Y Armfield; V R Dowell; D J Reinhardt
Journal:  J Clin Microbiol       Date:  1978-07       Impact factor: 5.948

2.  Recovery of anaerobic, facultative, and aerobic bacteria from clinical specimens in three anaerobic transport systems.

Authors:  A G Helstad; J L Kimball; D G Maki
Journal:  J Clin Microbiol       Date:  1977-06       Impact factor: 5.948

3.  A transport method for swab specimens submitted for aerobic and anaerobic bacteriology.

Authors:  M Alfa; A Lee
Journal:  J Clin Pathol       Date:  1982-05       Impact factor: 3.411

4.  Collection and transport of specimens for anaerobic culture.

Authors:  H O Hallander; A Flodström; C Aberg
Journal:  Infection       Date:  1980       Impact factor: 3.553

5.  Prospective, randomized, double-blind comparison of metronidazole and tobramycin with clindamycin and tobramycin in the treatment of intra-abdominal sepsis.

Authors:  J A Smith; A G Skidmore; A D Forward; A M Clarke; E Sutherland
Journal:  Ann Surg       Date:  1980-08       Impact factor: 12.969

6.  Anaerobic specimen transport system.

Authors:  W S Ramsey; W Hertl
Journal:  Experientia       Date:  1978-07-15

7.  Effects of storage in an anaerobic transport system on bacteria in known polymicrobial mixtures and in clinical specimens.

Authors:  G B Hill
Journal:  J Clin Microbiol       Date:  1978-12       Impact factor: 5.948

  7 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.