| Literature DB >> 11030993 |
Y Lin1, S Ahn, N Murali, W Brey, C R Bowers, W S Warren.
Abstract
Resistive or hybrid magnets can achieve substantially higher fields than those available in superconducting magnets, but their spatial homogeneity and temporal stability are unacceptable for high-resolution NMR. We show that modern stabilization and shimming technology, combined with detection of intermolecular zero-quantum coherences (iZQCs), can remove almost all of the effects of inhomogeneity and drifts, while retaining chemical shift differences and J couplings. In a 25-T electromagnet (1 kHz/s drift, 3 kHz linewidth over 1 cm(3)), iZQC detection removes >99% of the remaining inhomogeneity, to generate the first high-resolution liquid-state NMR spectra acquired at >1 GHz.Mesh:
Year: 2000 PMID: 11030993 DOI: 10.1103/PhysRevLett.85.3732
Source DB: PubMed Journal: Phys Rev Lett ISSN: 0031-9007 Impact factor: 9.161