Literature DB >> 11020747

Evaluation of a cerebral oximeter as a monitor of cerebral ischemia during carotid endarterectomy.

S K Samra1, E A Dy, K Welch, P Dorje, G B Zelenock, J C Stanley.   

Abstract

BACKGROUND: Stroke is an important contributor to perioperative morbidity and mortality associated with carotid endarterectomy (CEA). This investigation was designed to compare the performance of the INVOS-3100 cerebral oximeter to neurologic function, as a means of detecting cerebral ischemia induced by carotid cross-clamping, in patients undergoing carotid endarterectomy with cervical plexus block.
METHODS: Ninety-nine patients undergoing 100 CEAs with regional anesthesia (deep or superficial cervical plexus block) were studied. Bilateral regional cerebrovascular oxygen saturation (rSO2) was monitored using the INVOS-3100 cerebral oximeter. Patients were retrospectively assigned to one of two groups: those in whom a change in mental status or contralateral motor deficit was noted after internal carotid clamping (neurologic symptoms; n = 10) and those who did not show any neurologic change (no neurologic symptoms; n = 90). Data from 94 operations (neurologic symptoms = 10 and no neurologic symptoms = 84) were adequate for statistical analyses for group comparisons. A relative decrease in ipsilateral rSO2 after carotid occlusion (calculated as a percentage of preocclusion value) during all operations (n = 100) was also calculated to determine the critical level of rSO2 decrease associated with a change in neurologic function.
RESULTS: The mean (+/- SD) decrease in rSO2 after carotid occlusion in the neurologic symptoms group (from 63.2 +/- 8.4% to 51.0 +/- 11.6%) was significantly greater (P = 0.0002) than in the no neurologic symptoms group (from 65.8 +/- 8.5% to 61.0 +/- 9.3%). Logistic regression analysis used to determine if a change in rSO2, calculated as a percentage of preclamp value, could be used to predict change in neurologic function was highly significant (likelihood ratio chi-square = 13.7; P = 0.0002). A 20% decrease in rSO2 reading from the preclamp baseline, as a predictor of neurologic compromise, resulted in a sensitivity of 80% and specificity of 82.2%. The false-positive rate using this cutoff point was 66.7%, and the false-negative rate was 2.6%, providing a positive predictive value of 33.3% and a negative predictive value of 97.4%.
CONCLUSION: Monitoring rSO2 with INVOS-3100 to detect cerebral ischemia during CEA has a high negative predictive value, but the positive predictive value is low.

Entities:  

Mesh:

Substances:

Year:  2000        PMID: 11020747     DOI: 10.1097/00000542-200010000-00015

Source DB:  PubMed          Journal:  Anesthesiology        ISSN: 0003-3022            Impact factor:   7.892


  47 in total

1.  Comparison of two spatially resolved NIRS oxygenation indices.

Authors:  Roy E Gagnon; Andrew J Macnab; Faith A Gagnon; Derek Blackstock; Jacques G LeBlanc
Journal:  J Clin Monit Comput       Date:  2002-12       Impact factor: 2.502

Review 2.  Haemoglobin oxygen saturation as a biomarker: the problem and a solution.

Authors:  David A Boas; Maria Angela Franceschini
Journal:  Philos Trans A Math Phys Eng Sci       Date:  2011-11-28       Impact factor: 4.226

3.  Different ventilation techniques and hemodynamic optimization to maintain regional cerebral oxygen saturation (rScO2) during laparoscopic bariatric surgery: a prospective randomized interventional study.

Authors:  Osama M Asaad
Journal:  J Anesth       Date:  2018-04-06       Impact factor: 2.078

4.  Cerebral oxygenation and processed EEG response to clamping and shunting during carotid endarterectomy under general anesthesia.

Authors:  William Perez; Christopher Dukatz; Sami El-Dalati; James Duncan; Mahmoud Abdel-Rasoul; Andrew Springer; Michael R Go; Roger Dzwonczyk
Journal:  J Clin Monit Comput       Date:  2015-01-09       Impact factor: 2.502

5.  Comparing the response of pulse oximetry and regional cerebral oxygen saturation to hypoxia in preschool children.

Authors:  Yi Lu; Meiqin Di; Chan Li; Mengmeng Chen; Kaiming Yuan; Wangning Shangguan
Journal:  Exp Ther Med       Date:  2019-11-14       Impact factor: 2.447

6.  Foreseeing the danger in the beach chair position: Are standard measurement methods reliable?

Authors:  Baris Kocaoglu; Serpil Ustalar Ozgen; Fevzi Toraman; Mustafa Karahan; Osman Guven
Journal:  Knee Surg Sports Traumatol Arthrosc       Date:  2014-06-07       Impact factor: 4.342

7.  Assessment of cerebral oxygenation using near infrared spectroscopy during isovolemic hemodilution in pediatric patients.

Authors:  Joseph D Tobias
Journal:  J Clin Monit Comput       Date:  2011-08-26       Impact factor: 2.502

8.  Quantitative analysis of the effect of end-tidal carbon dioxide on regional cerebral oxygen saturation in patients undergoing carotid endarterectomy under general anaesthesia.

Authors:  Seung-Hee Ki; Jin-Ho Rhim; Jae-Hong Park; Young-Jin Han; Yong-Pil Cho; Tae-Won Kwon; Byung-Moon Choi; Gyu-Jeong Noh
Journal:  Br J Clin Pharmacol       Date:  2017-10-26       Impact factor: 4.335

9.  Measurements of oxygen saturation of brain, liver and heart areas in the supine and sitting position using near infrared spectrophotometry.

Authors:  Athanasia Tsaroucha; Anteia Paraskeva; Argyro Fassoulaki
Journal:  Rom J Anaesth Intensive Care       Date:  2017-10

10.  Bilateral near infrared spectroscopy in space-occupying middle cerebral artery stroke.

Authors:  M S Damian; R Schlosser
Journal:  Neurocrit Care       Date:  2007       Impact factor: 3.210

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.