Suppression of intravascular signal on fat-saturated contrast-enhanced thoracic MR arteriograms.

E S Siegelman1, R Charafeddine, A H Stolpen, L Axel.   

Abstract

PURPOSE: To assess the prevalence of artifactual signal intensity loss within the aortic arch and proximal branch vessels on fat-saturated contrast material-enhanced magnetic resonance (MR) arteriograms of the thoracic aorta and to hypothesize about the cause of the loss of signal intensity.
MATERIALS AND METHODS: Between January and June 1998, 105 consecutive MR arteriograms of the thoracic aorta were acquired in 103 patients at 1.5 T. Imaging included an arterial phase three-dimensional (3D) fat-saturated contrast-enhanced gradient-echo (GRE) sequence followed by a delayed two-dimensional (2D) transverse fat-saturated GRE sequence. All MR images were reviewed by two radiologists who were blinded to patient history and results of imaging studies and who evaluated the images for the presence of intraluminal loss of signal intensity in the aortic arch and the proximal branch vessels.
RESULTS: Intravascular loss of signal intensity was present in at least one vessel on 23 of the 105 arterial phase 3D studies. Seventy-one of 91 left subclavian arterial segments had loss of signal intensity on the delayed 2D studies.
CONCLUSION: Intravascular signal intensity loss can be present on contrast-enhanced fat-saturated images of the aortic arch and proximal branch vessels, particularly the left subclavian artery. This phenomenon, which is to the authors' knowledge previously unreported and which is hypothesized to result from undesired water saturation, should not be misinterpreted as stenotic or occlusive vascular disease.

Entities:  

Mesh:

Substances:

Year:  2000        PMID: 11012431     DOI: 10.1148/radiology.217.1.r00oc21115

Source DB:  PubMed          Journal:  Radiology        ISSN: 0033-8419            Impact factor:   11.105


  3 in total

1.  Pseudostenosis of the internal carotid artery in 3D time-of-flight MR angiography: effects of a magnetization transfer contrast pulse and metallic material.

Authors:  Shinji Naganawa; Tokiko Koshikawa; Kimihide Sato; Toshio Katagiri; Takeo Mimura; Takeo Ishigaki; Ikuo Aoki
Journal:  Eur Radiol       Date:  2003-04-10       Impact factor: 5.315

2.  Bilateral breast MRI by use of dual-source parallel radiofrequency excitation and image-based shimming at 3 Tesla: improvement in homogeneity on fat-suppression imaging.

Authors:  Kinya Ishizaka; Fumi Kato; Satoshi Terae; Suzuko Mito; Noriko Oyama-Manabe; Tamotsu Kamishima; Mitsuhiro Nakanishi; Hiroyuki Sugimori; Hiroyuki Hamaguchi; Hiroki Shirato
Journal:  Radiol Phys Technol       Date:  2014-06-26

3.  Gadolinium susceptibility artifact causing false positive stenosis isolated to the proximal common carotid artery in 3D dynamic contrast medium enhanced MR angiography of the thorax--a brief review of causes and prevention.

Authors:  A Temel Tirkes; Mark A Rosen; Evan S Siegelman
Journal:  Int J Cardiovasc Imaging       Date:  2003-04       Impact factor: 2.357

  3 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.