Lens of the eye: radiation dose in balloon dacryocystoplasty.

E T Ilgit1, N Meric, D Bor, I Oznur, O Konus, S Isik.   

Abstract

PURPOSE: To evaluate the absorbed radiation dose to the lens of the eye, which is the critical organ in the primary beam during fluoroscopically guided transluminal balloon dilation of the lacrimal drainage system (balloon dacryocystoplasty) for obstructive epiphora and to evaluate the possibility of deterministic radiation effect on the lens.
MATERIALS AND METHODS: The radiation dose to the lens of the eye during balloon dacryocystoplasty (which includes pre- and postintervention dacryocystography) was measured in 10 consecutive patients by using thermoluminescent dosimeters on the lids of both eyes as close as possible to the lenses. A C-arm angiographic unit coupled with a digital imaging system was used, with similar exposure and geometric parameters in all cases.
RESULTS: The mean radiation dose to the lens of the treated eye was 4.6 mGy +/- 2.2 (dose range, 1.9-9.1 mGy) and to that of the untreated eye was 38.5 mGy +/- 17.5 (dose range, 14.7-67.8 mGy).
CONCLUSION: The lens of the untreated eye receives a higher dose than that of the treated eye because of its closer proximity to the x-ray tube in a lateral projection. In the lens, even the highest measured radiation dose (67.8 mGy) still was well below the deterministic threshold for lens opacity and cataract formation.

Entities:  

Mesh:

Year:  2000        PMID: 11012423     DOI: 10.1148/radiology.217.1.r00oc0454

Source DB:  PubMed          Journal:  Radiology        ISSN: 0033-8419            Impact factor:   11.105


  5 in total

1.  Radiation exposure of operator performing interventional procedures using a flat panel angiography system: evaluation with photoluminescence glass dosimeters.

Authors:  Yoshinori Funama; Nozomu Nagasue; Kazuo Awai; Ichiro Sakamoto; Kiyotaka Kakei; Masamichi Shimamura; Yasuyuki Yamashita; Masataka Uetani
Journal:  Jpn J Radiol       Date:  2010-07-27       Impact factor: 2.374

2.  [Balloon dilatation and Stentimplantation of the nasolacrimal duct for chronic epiphora].

Authors:  U Lachmund; D Ammann-Rauch; A Forrer; M Grob; C Petralli; L Remonda; T Roeren; K Wilhelm
Journal:  Ophthalmologe       Date:  2005-04       Impact factor: 1.059

3.  [Balloon dilatation of the canaliculus communis for chronic epiphora].

Authors:  U Lachmund; D Ammann; A Forrer; C Petralli; L Remonda; T Roeren; F Vonmoos; K Wilhelm
Journal:  Ophthalmologe       Date:  2005-04       Impact factor: 1.059

4.  Real-time eye lens dose monitoring during cerebral angiography procedures.

Authors:  M J Safari; J H D Wong; K A A Kadir; N K Thorpe; D L Cutajar; M Petasecca; M L F Lerch; A B Rosenfeld; K H Ng
Journal:  Eur Radiol       Date:  2015-05-23       Impact factor: 5.315

5.  MR dacryocystography in the evaluation of patients with obstructive epiphora treated by means of interventional radiologic procedures.

Authors:  B Coskun; E Ilgit; B Onal; O Konuk; G Erbas
Journal:  AJNR Am J Neuroradiol       Date:  2011-12-15       Impact factor: 3.825

  5 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.