Literature DB >> 10946194

Astigmatism correction with a foldable toric intraocular lens in cataract patients.

I Ruhswurm1, U Scholz, M Zehetmayer, G Hanselmayer, C Vass, C Skorpik.   

Abstract

PURPOSE: To determine the efficacy and rotational stability of a toric posterior chamber silicone intraocular lens (IOL) to correct preoperative astigmatism in cataract patients.
SETTING: Department of Ophthalmology, University Hospital of Vienna, Vienna Austria.
METHODS: Between 1993 and 1998, foldable toric single-piece plate-haptic silicone IOLs were implanted in 37 eyes (30 patients). The cylindrical IOL power was 2.00 diopters (D) (n = 29), 3.50 D (n = 7), or 4.00 D (n = 1). Phacoemulsification was performed through a scleral or a corneal sutureless self-sealing incision. Outcomes of Snellen visual acuity (without, with spherical, and with best correction), refractive and keratometric astigmatism, and IOL rotation after early postoperative (mean 15.9 days +/- 10.1 [SD]) and long-term (mean 20.3 +/- 16.6 months) follow-ups were evaluated.
RESULTS: At last follow-up, 31 eyes (83.8%) had a spherically corrected and 34 (91.9%) a best corrected visual acuity of 0.5 (20/40) or better. Mean preoperative refractive and keratometric astigmatism was 2.68 and 2.70 D, respectively. At the last postoperative follow-up, mean refractive astigmatism was reduced to 0.84 D; keratometric astigmatism was 2.30 D. In 7 eyes (18.9%), the IOL axis was rotated a maximum of 25 degrees. In all 37 eyes, the axis of the toric IOL remained within 30 degrees of rotation.
CONCLUSIONS: Early postoperative and long-term follow-ups showed effective and stable correction of astigmatism after implantation of a foldable toric posterior chamber silicone IOL.

Entities:  

Mesh:

Substances:

Year:  2000        PMID: 10946194     DOI: 10.1016/s0886-3350(00)00317-5

Source DB:  PubMed          Journal:  J Cataract Refract Surg        ISSN: 0886-3350            Impact factor:   3.351


  31 in total

Review 1.  [Intraocular lenses for the correction of refraction errors. Part II. Phakic posterior chamber lenses and refractive lens exchange with posterior chamber lens implantation].

Authors:  T Kohnen; T Kasper; E Terzi
Journal:  Ophthalmologe       Date:  2005-11       Impact factor: 1.059

Review 2.  [Intraocular lenses for the correction of refraction errors. Part 1: phakic anterior chamber lenses].

Authors:  T Kohnen; M Baumeister; M Cichocki
Journal:  Ophthalmologe       Date:  2005-10       Impact factor: 1.059

3.  [Determination of toric intraocular lenses].

Authors:  A Langenbucher; Anja Viestenz; N Szentmáry; W Behrens-Baumann; A Viestenz
Journal:  Ophthalmologe       Date:  2008-07       Impact factor: 1.059

4.  Effects of the toric intraocular lens on correction of preexisting corneal astigmatism.

Authors:  Hiroshi Sasaki; Motoaki Yoshida; Shin-Ichi Manabe; Koichi Yoshimura; Ken Hayashi
Journal:  Jpn J Ophthalmol       Date:  2012-08-08       Impact factor: 2.447

5.  [Correction of corneal astigmatism with toric lenses : Theory and clinical aspects].

Authors:  A Langenbucher; A Viestenz; N Szentmáry; A Viestenz; T Eppig; B Seitz
Journal:  Ophthalmologe       Date:  2010-02       Impact factor: 1.059

6.  Simple and accurate alignment of toric intraocular lenses and evaluation of their rotation errors using anterior segment optical coherence tomography.

Authors:  Kazuhiro Watanabe; Kazuno Negishi; Hidemasa Torii; Megumi Saiki; Murat Dogru; Kazuo Tsubota
Journal:  Jpn J Ophthalmol       Date:  2011-10-29       Impact factor: 2.447

7.  Combined 23-gauge microincisonal vitrectomy surgery and phacoemulsification with AcrySof toric intraocular lens implantation: a comparative study.

Authors:  D H Park; J P Shin; S Y Kim
Journal:  Eye (Lond)       Date:  2011-07-15       Impact factor: 3.775

8.  Comparison of the keratometric corneal astigmatism and refractive astigmatism after phacoemulsification and foldable intraocular lens implantation.

Authors:  Ali Sharifi; Leila Sharifi; Afsaneh Morteza
Journal:  Int Ophthalmol       Date:  2012-06-09       Impact factor: 2.031

9.  [Toric intraocular lenses for correction of persistent corneal astigmatism after penetrating keratoplasty].

Authors:  A Viestenz; M Küchle; B Seitz; A Langenbucher
Journal:  Ophthalmologe       Date:  2005-02       Impact factor: 1.059

Review 10.  [Toric IOLs after cataract surgery and refractive lens exchange].

Authors:  G U Auffarth; T M Rabsilber
Journal:  Ophthalmologe       Date:  2007-12       Impact factor: 1.059

View more