Influence of infusion line compliance on drug delivery rate during acute line loop formation.

M Weiss1, O Bänziger, T Neff, S Fanconi.   

Abstract

OBJECTIVE: To determine whether infusion line compliance contributes to irregular drug delivery during vertical displacement of syringe pumps.
DESIGN: Five different commercially available infusion lines were studied at infusion rates of 0.5, 1.0, and 1.5 ml/h. Zero drug delivery time was measured after acute line loop formation (70 cm) using an electronic balance. Compliance of each infusion line was calculated using a pressure transducer and measurement of the occlusion release bolus at 300 mmHg occlusion pressure. Finally, the influence of infusion line compliance on drug delivery during acute lowering of the syringe pump was studied using low- and high-compliance infusion lines.
RESULTS: Acute line loop formation resulted in zero drug delivery time from 5.1 +/- 1.5 to 44.0 +/- 6.8 s at flow rates of 0.5 ml/h. Increased flow rates significantly reduced loop-induced flow variability. A close correlation was found between zero drug delivery time and calculated infusion line compliance at 0.5 ml/h (linear regression R2 = 0.79). Lowering of the syringe pump 50 cm prolonged zero drug delivery time from 295.8 +/- 20.7 s with the low-compliance tube to 463.3 +/- 24.0 s with the high-compliance infusion line.
CONCLUSIONS: Infusion line compliance contributes to irregular drug delivery associated with vertical displacement of syringe pumps. Siphoning of the infusion line during patient care should be avoided, and flow rates of 1 ml/h or higher are recommended. Low-compliance infusion lines are indicated whenever highly short-acting vasoactive drugs at low delivery rates are administered.

Entities:  

Mesh:

Year:  2000        PMID: 10945397     DOI: 10.1007/s001340051246

Source DB:  PubMed          Journal:  Intensive Care Med        ISSN: 0342-4642            Impact factor:   17.440


  5 in total

1.  In-line filter included into the syringe infusion pump assembly reduces flow irregularities.

Authors:  B Brotschi; B Grass; M Weiss; C Doell; V Bernet
Journal:  Intensive Care Med       Date:  2012-01-12       Impact factor: 17.440

2.  Delay and stability of central venous administration of norepinephrine in children: a bench study.

Authors:  Mehdi Oualha; Mikaël Capelo; Odile Spreux-Varoquaux; Isabelle Drouet-Chaillou; Jean-Marc Tréluyer; Philippe Hubert; Fabrice Lesage
Journal:  Eur J Pediatr       Date:  2013-08-11       Impact factor: 3.183

3.  Analytical method for calculation of deviations from intended dosages during multi-infusion.

Authors:  Maurits K Konings; Roland A Snijder; Joris H Radermacher; Annemoon M Timmerman
Journal:  Biomed Eng Online       Date:  2017-01-17       Impact factor: 2.819

4.  A Hazard Analysis of Class I Recalls of Infusion Pumps.

Authors:  Xuemei Gao; Qiang Wen; Xiaolian Duan; Wei Jin; Xiaohong Tang; Ling Zhong; Shitao Xia; Hailing Feng; Daidi Zhong
Journal:  JMIR Hum Factors       Date:  2019-05-03

5.  Changeovers of vasoactive drug infusion pumps: impact of a quality improvement program.

Authors:  Laurent Argaud; Martin Cour; Olivier Martin; Marc Saint-Denis; Tristan Ferry; Agnes Goyatton; Dominique Robert
Journal:  Crit Care       Date:  2007       Impact factor: 9.097

  5 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.