Literature DB >> 10938964

NAT and viral safety in blood transfusion.

W K Roth1, S Buhr, C Drosten, E Seifried.   

Abstract

BACKGROUND AND OBJECTIVES: Nucleic acid testing was introduced at our blood transfusion service in order to reduce the diagnostic preseroconversion window for transfusion relevant viruses.
MATERIALS AND METHODS: Up to 96 donor samples were pooled overnight by two Tecan Genesis RSP 150 pipettors. Pools were centrifuged at 48,000 g for one hour to enrich viruses. Viral nucleic acids were extracted from centrifugation pellets using modified Qiagen viral RNA kit. HIV and HBV sequences were amplified by in house TaqMan PCR's with patented primers and probes for HBV. HCV PCR was performed by Roche Amplicor V2.0. "Nadis" software was developed for pooling, resolving positive pools and data communication with main frame computer.
RESULTS: PCR testing was introduced in January 1997 for HCV HBV and HIV-1 for all plasma products and for labile components at the 21st of April 1997. Throughput increased from 2000 to 4000 samples per day. PCR testing is done in parallel to serological testing and products are released eight hours after pooling is completed. Until January 2000, 1,078,940 donations have been tested in 13,274 pools. A total of seven PCR only positives were identified (3 HCV, 3 HBV, 1 HIV).
CONCLUSION: The yield of PCR testing for transfusion relevant viruses confirms theoretical estimates on the residual risk of antibody screening.

Entities:  

Mesh:

Substances:

Year:  2000        PMID: 10938964

Source DB:  PubMed          Journal:  Vox Sang        ISSN: 0042-9007            Impact factor:   2.144


  7 in total

Review 1.  Safety of the blood supply in Latin America.

Authors:  Gabriel A Schmunis; Jose R Cruz
Journal:  Clin Microbiol Rev       Date:  2005-01       Impact factor: 26.132

Review 2.  Emerging Pathogens - How Safe is Blood?

Authors:  Michael Schmidt; Wolf-Jochen Geilenkeuser; Walid Sireis; Erhard Seifried; Kai Hourfar
Journal:  Transfus Med Hemother       Date:  2013-12-27       Impact factor: 3.747

3.  Evaluation of the COBAS amplicor HBV monitor assay and comparison with the ultrasensitive HBV hybrid capture 2 assay for quantification of hepatitis B virus DNA.

Authors:  Eric Q Konnick; Maria Erali; Edward R Ashwood; David R Hillyard
Journal:  J Clin Microbiol       Date:  2005-02       Impact factor: 5.948

4.  A study on optimization of plasma pool size for viral infectious markers in Indian blood donors using nucleic acid amplification testing.

Authors:  Ankit Mathur; Jwalant Shah; Ripal Shah; Priti Shah; V Harimoorthy; N Choudhury; Sunita Tulsiani
Journal:  Asian J Transfus Sci       Date:  2012-01

5.  Sufficient blood, safe blood: can we have both?

Authors:  Halvard Bönig; Michael Schmidt; Kai Hourfar; Jörg Schüttrumpf; Erhard Seifried
Journal:  BMC Med       Date:  2012-03-23       Impact factor: 8.775

6.  An Internal Reference Control Duplex Real-Time Polymerase Chain Reaction Assay for Detecting Bacterial Contamination in Blood Products.

Authors:  Jin-Ju Zhang; Jing-Jing Tian; Shuang-Shi Wei; Sheng-Bao Duan; Hong-Mei Wang; Ye-Zhou Chen; Shao-Hua Ding; Chun Zhang; Qing-Lin Meng; Yong Li
Journal:  PLoS One       Date:  2015-07-31       Impact factor: 3.240

7.  Novel multiple swab method enables high efficiency in SARS-CoV-2 screenings without loss of sensitivity for screening of a complete population.

Authors:  Michael Schmidt; Sebastian Hoehl; Annemarie Berger; Heinz Zeichhardt; Kai Hourfar; Sandra Ciesek; Erhard Seifried
Journal:  Transfusion       Date:  2020-07-23       Impact factor: 3.337

  7 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.