Literature DB >> 10931571

MR evaluation ex vivo and in vivo of a covered stent-graft for abdominal aortic aneurysms: ferromagnetism, heating, artifacts, and velocity mapping.

L Engellau1, J Olsrud, S Brockstedt, U Albrechtsson, L Norgren, F Ståhlberg, E M Larsson.   

Abstract

Magnetic resonance imaging (MRI) safety was evaluated at 1.5 T in a covered nickel titanium stent-graft (Vanguard) used for endovascular treatment of abdominal aortic aneurysms (AAAs). Imaging artifacts were assessed on MRI with contrast-enhanced (CE) three-dimensional (3D) MR angiography (MRA) and spiral computed tomography (CT) in 10 patients as well as ex vivo. Velocity mapping was performed in the suprarenal aorta and femoral arteries in 14 patients before and after stent-graft placement. For comparison it was also performed in six healthy volunteers. No ferromagnetism or heating was detected. Metal artifacts caused minimal image distortion on MRI/MRA. The artifacts disturbed image evaluation on CT at the graft bifurcation and graft limb junction. No significant differences in mean flow were found in patients before and after stent-graft placement. Our study indicates that MRI at 1.5 T may be performed safely in patients with the (Vanguard) stent-graft. MRI/MRA provides diagnostic image information. Velocity mapping is not included in our routine protocol.

Entities:  

Mesh:

Substances:

Year:  2000        PMID: 10931571     DOI: 10.1002/1522-2586(200007)12:1<112::aid-jmri13>3.0.co;2-g

Source DB:  PubMed          Journal:  J Magn Reson Imaging        ISSN: 1053-1807            Impact factor:   4.813


  6 in total

1.  MR-velocity mapping in vascular stents to assess peak systolic velocity. In vitro comparison of various stent designs made of Stainless Steel and Nitinol.

Authors:  Jacqueline van Holten; Patrik Kunz; Paul G H Mulder; Peter M T Pattynama; Hildo J Lamb; Lukas C van Dijk
Journal:  MAGMA       Date:  2002-11       Impact factor: 2.310

2.  Thoracic aortic stent graft: comparison of contrast-enhanced MR angiography and CT angiography in the follow-up: initial results.

Authors:  Stefanie Weigel; Bernd Tombach; David Maintz; Stefan Klotz; Thomas Vestring; Walter Heindel; Roman Fischbach
Journal:  Eur Radiol       Date:  2003-02-19       Impact factor: 5.315

3.  Safety of cardiovascular magnetic resonance in patients with cardiovascular implants and devices.

Authors:  S K Prasad; D J Pennell
Journal:  Heart       Date:  2004-11       Impact factor: 5.994

Review 4.  MRI in guiding and assessing intramyocardial therapy.

Authors:  M Saeed; D Saloner; O Weber; A Martin; C Henk; C Higgins
Journal:  Eur Radiol       Date:  2005-01-29       Impact factor: 5.315

5.  MRI assessment of thoracic stent grafts after emergency implantation in multi trauma patients: a feasibility study.

Authors:  Volker Rasche; Alexander Oberhuber; Stephan Trumpp; Axel Bornstedt; Karl-Heinz Orend; Nico Merkle; Wolfgang Rottbauer; Martin Hoffmann
Journal:  Eur Radiol       Date:  2011-02-18       Impact factor: 5.315

Review 6.  An overview on the advances in cardiovascular interventional MR imaging.

Authors:  Olaf Saborowski; Maythem Saeed
Journal:  MAGMA       Date:  2007-05-09       Impact factor: 2.310

  6 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.