Literature DB >> 10928531

Use of fibrin sealant as a hemostatic agent after liver biopsy in swine.

E K Paulson1, G R Stephenson, M C Neal, V Rossin, J H Lawson.   

Abstract

PURPOSE: To determine whether fibrin sealant injected into the tract created by liver biopsy can be used to decrease postprocedural bleeding. An innovative delivery system was used to deploy the fibrin sealant.
MATERIALS AND METHODS: Fibrin sealant is a hemostatic agent consisting of a suspension of fibrinogen and thrombin. A delivery system was devised whereby fibrin sealant could be injected into the tract created by liver biopsy. Thirty swine were randomized into three groups: control (n = 10), heparin (n = 10), and warfarin (n = 10). Each swine underwent laparotomy and was randomized to undergo three to five open liver biopsies with either a 14-gauge cutting needle in conjunction with the fibrin sealant device or a standard 14-gauge cutting needle alone. Forty-seven biopsy procedures were performed with the device; 64 biopsy procedures were performed without the device. Immediate blood loss per biopsy (mL) was estimated based on the size of the blood stain on a sponge. Specimens were assessed for sample size.
RESULTS: Immediate blood loss with and without the device, respectively, was: control, 0.1 mL, 5.4 mL; heparin, 0 mL, 7 mL; warfarin, 0.1 mL, 9.3 mL. These differences were significant (P < .01) for each group of swine. In 43 of 47 biopsies (91%), the device functioned without difficulty. There was no difference in sample size when the device was used.
CONCLUSIONS: The fibrin sealant device is effective in reducing bleeding after open liver biopsy in anticoagulated and nonanticoagulated swine. The promising results suggest that a trial of percutaneous liver biopsy in swine should be considered.

Entities:  

Mesh:

Substances:

Year:  2000        PMID: 10928531     DOI: 10.1016/s1051-0443(07)61810-7

Source DB:  PubMed          Journal:  J Vasc Interv Radiol        ISSN: 1051-0443            Impact factor:   3.464


  4 in total

1.  Radiofrequency cauterization with biopsy introducer needle.

Authors:  William F Pritchard; Diane Wray-Cahen; John W Karanian; Stephen Hilbert; Bradford J Wood
Journal:  J Vasc Interv Radiol       Date:  2004-02       Impact factor: 3.464

2.  New Radiofrequency Device to Reduce Bleeding after Core Needle Biopsy: Experimental Study in a Porcine Liver Model.

Authors:  Sanghyeok Lim; Hyunchul Rhim; Min Woo Lee; Kyoung Doo Song; Tae Wook Kang; Young-Sun Kim; Hyo Keun Lim
Journal:  Korean J Radiol       Date:  2017-01-05       Impact factor: 3.500

3.  Injecting hemostatic matrix in the path of biopsies: efficacy, potential complications, and the management of such complications.

Authors:  Antonio Rahal Junior; Priscila Mina Falsarella; Vinicius Tadeu Rodrigues Ferreira; Guilherme Cayres Mariotti; Marcos Roberto Gomes de Queiroz; Rodrigo Gobbo Garcia
Journal:  Radiol Bras       Date:  2018 Mar-Apr

4.  Gelatin-thrombin hemostatic matrix injection to salvage refractory post-renal graft biopsy bleed.

Authors:  V Jain; A Gupta; A Gulia; M Singhal; S Gulati; S C Tiwari; A Kumar
Journal:  Indian J Nephrol       Date:  2013-03