Evaluation of automated nucleic acid extraction devices for application in HCV NAT.

J M Jongerius1, M Bovenhorst, C L van der Poel, J A van Hilten, A C Kroes, J A van der Does, E F van Leeuwen, R Schuurman.   

Abstract

BACKGROUND: To further improve the safety of the blood supply, various national blood transfusion organizations presently use or are in the process of implementing routine HCV NAT in minipools. According to the Committee for Proprietary Medicinal Products (CPMP) of the European Union, the HCV NAT detection limit of the assay should be 100 IU per mL (270 geq/mL) for testing initial plasma pools. Paul Ehrlich Institute (PEI) regulations stipulate that 5000 IU per mL (13,500 geq/mL) must be detected to calculate the amount contributed by individual donations composing the minipool. The sensitivity for HCV RNA extraction achieved by three commercially available laboratory kits was compared. STUDY DESIGN AND METHODS: Nucleic acids from 1-in-3 serial dilutions of an HCV RNA run control (Pelispy, CLB) were extracted with three kits (Cobas Amplicor, Roche Diagnostic Systems; BioRobot 9604, Qiagen; and NucliSens Extractor, Organon Teknika). HCV PCR of all extracts was performed using a second-generation Cobas Amplicor HCV test and the Cobas Amplicor analyzer.
RESULTS: The manual Cobas Amplicor, the BioRobot 9604, and the NucliSens Extractor setups allow a 95-percent HCV RNA detection limit of 129, 82, and 12 geq per mL, respectively. The maximal pool size for the manual Cobas Amplicor, the BioRobot 9604, and the NucliSens Extractor kits that would still meet the PEI criteria for HCV NAT in minipools was calculated at 104, 164, and 1125 donations, respectively.
CONCLUSION: All three HCV NAT kits evaluated meet the criteria set by CPMP and PEI. The highest sensitivity for HCV NAT screening can be achieved with the high-volume NucliSens Extractor method in combination with the Cobas Amplicor HCV v2.0 test on the Cobas Amplicor analyzer.

Entities:  

Mesh:

Substances:

Year:  2000        PMID: 10924618     DOI: 10.1046/j.1537-2995.2000.40070871.x

Source DB:  PubMed          Journal:  Transfusion        ISSN: 0041-1132            Impact factor:   3.157


  3 in total

1.  Automated procedure for improving the RNA isolation step in viral load testing for human immunodeficiency virus.

Authors:  Min Xu; Yung Chan; Steven H Fischer; Alan T Remaley
Journal:  J Clin Microbiol       Date:  2004-01       Impact factor: 5.948

2.  Comparison of automated and manual nucleic acid extraction methods for detection of enterovirus RNA.

Authors:  Julia H Knepp; Melissa A Geahr; Michael S Forman; Alexandra Valsamakis
Journal:  J Clin Microbiol       Date:  2003-08       Impact factor: 5.948

3.  Efficient extraction of virus DNA by NucliSens Extractor allows sensitive detection of hepatitis B virus by PCR.

Authors:  E Gobbers; T A Oosterlaken; M J van Bussel; R Melsert; A C Kroes; E C Claas
Journal:  J Clin Microbiol       Date:  2001-12       Impact factor: 5.948

  3 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.