Performance of three blood glucose meters.

C C Rheney1, J K Kirk.   

Abstract

OBJECTIVE: To evaluate the performance of three blood glucose meters.
METHODS: The One Touch II (LifeScan, Milpitas, CA), Glucometer Elite (Bayer, Elkhart, IN), and Accu-Chek Advantage (Boehringer Mannheim, Indianapolis, IN) were compared with a reference laboratory method (Technicon Chem System, Tarrytown, NY). Blood glucose meters used in this study were validated by a clinically oriented approach known as the error grid analysis (EGA), for which the performance of the meters was compared to a laboratory standard, and by the criteria of the American Diabetes Association (ADA). Limits of agreement were evaluated using differences from the reference laboratory method and 95% Cls. Capillary blood was obtained from study participants in fasting state with the morning blood draw and tested on the three meters simultaneously.
RESULTS: A total of 120 blood glucose meter readings were analyzed; values ranged from 62 to 396 mg/dL. For all three meters, at least 75% of the capillary blood glucose values fell into zone A (acceptable) of the EGA. The number of values falling into zone B (unacceptable) were 10, 8, and 6 for the Accu-Chek Advantage, the One Touch II, and the Glucometer Elite, respectively. Only 15-25% of the meter glucose readings met the ADA criteria of being within 5% of the laboratory standard. The mean difference from the reference values was least with the Glucometer Elite.
CONCLUSIONS: The majority of blood glucose determinations obtained on the meters used in this study were within the acceptable limits using the EGA. The Glucometer Elite meter had the fewest values in the unacceptable range and had the least mean difference from reference laboratory values.

Entities:  

Mesh:

Year:  2000        PMID: 10917376     DOI: 10.1345/aph.19187

Source DB:  PubMed          Journal:  Ann Pharmacother        ISSN: 1060-0280            Impact factor:   3.154


  8 in total

1.  Phytochemical and biological studies of Adiantum capillus-veneris L.

Authors:  Zedan Z Ibraheim; Amany S Ahmed; Yaser G Gouda
Journal:  Saudi Pharm J       Date:  2011-01-18       Impact factor: 4.330

2.  Accuracy of the modified Continuous Glucose Monitoring System (CGMS) sensor in an outpatient setting: results from a diabetes research in children network (DirecNet) study.

Authors:  Michael J Tansey; Roy W Beck; Bruce A Buckingham; Nelly Mauras; Rosanna Fiallo-Scharer; Dongyuan Xing; Craig Killman; William V Tamborlane; Katrina J Ruedy
Journal:  Diabetes Technol Ther       Date:  2005-02       Impact factor: 6.118

3.  Quality assessment of patients' self-monitoring of blood glucose in community pharmacies.

Authors:  Reidun L S Kjome; Anne G Granas; Kari Nerhus; Sverre Sandberg
Journal:  Pharm Pract (Granada)       Date:  2010-03-15

4.  Frontal lobe regulation of blood glucose levels: support for the limited capacity model in hostile violence-prone men.

Authors:  Robert P Walters; Patti Kelly Harrison; Ransom W Campbell; David W Harrison
Journal:  Brain Inform       Date:  2016-02-01

5.  High prevalence of cardiovascular risk factors in Peruvian adolescents living in a peri-urban shantytown: a cross-sectional study.

Authors:  Elizabeth S Abbs; José Viñoles; Jorge O Alarcón; Heather M Johnson; Joseph R Zunt
Journal:  J Health Popul Nutr       Date:  2017-05-22       Impact factor: 2.000

6.  Divergent fifteen-year trends in traditional and cardiometabolic risk factors of cardiovascular diseases in the Seychelles.

Authors:  Pascal Bovet; Sarah Romain; Conrad Shamlaye; Shanti Mendis; Roger Darioli; Walter Riesen; Luc Tappy; Fred Paccaud
Journal:  Cardiovasc Diabetol       Date:  2009-06-26       Impact factor: 9.951

7.  Diabetes and pre-diabetes are associated with cardiovascular risk factors and carotid/femoral intima-media thickness independently of markers of insulin resistance and adiposity.

Authors:  David Faeh; Julita William; Patrick Yerly; Fred Paccaud; Pascal Bovet
Journal:  Cardiovasc Diabetol       Date:  2007-10-24       Impact factor: 9.951

8.  Prevalence, awareness and control of diabetes in the Seychelles and relationship with excess body weight.

Authors:  David Faeh; Julita William; Luc Tappy; Eric Ravussin; Pascal Bovet
Journal:  BMC Public Health       Date:  2007-07-19       Impact factor: 3.295

  8 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.