Literature DB >> 10895006

The artificial urinary sphincter in 86 patients with intrinsic sphincter deficiency: satisfactory actuarial adequate function rates.

J L Bosch1, A J Klijn, F H Schröder, W C Hop.   

Abstract

OBJECTIVE: To evaluate the long-term results of the AMS-800 artificial urinary sphincter (AUS) in patients with incontinence due to intrinsic sphincter deficiency, taking into account the continence status and the durability of the device. PATIENTS AND METHODS: Between 1984 and 1997, an AUS was implanted in 86 patients. Kaplan-Meier survival analysis was used to determine the 'primary adequate function' rate (P-AF) and the 'additional procedure assisted adequate function' rate (APA-AF). These rates define adequate function as satisfactory continence (use of maximally 1 pad per 24 h) in combination with good AUS function. In the case of P-AF, no revisions have been required. In the case of APA-AF, one or more revisions of parts of the AUS are acceptable but a complete exchange or explantation of the AUS marks its endpoint.
RESULTS: At last follow-up, satisfactory continence was found in 76% of the patients. Continence was markedly improved in another 7%. The 5-year P-AF and APA-AF rates were 46 and 67%, respectively. The 5-year P-AF rates for the periods before and after the introduction of the narrow backed cuff were 33 and 61%, respectively (p = 0.03).
CONCLUSIONS: The AUS can give excellent results as far as urinary continence is concerned, but only at the expense of a considerable reoperation rate. The 5-year actuarial primary adequate function rate has almost doubled since the introduction of the narrow backed cuff design.

Entities:  

Mesh:

Year:  2000        PMID: 10895006     DOI: 10.1159/000020273

Source DB:  PubMed          Journal:  Eur Urol        ISSN: 0302-2838            Impact factor:   20.096


  4 in total

1.  Treating stress urinary incontinence in female patients with neuropathic bladder: the value of the autologous fascia rectus sling.

Authors:  Anastasios Athanasopoulos; Konstantinos Gyftopoulos; Edward J McGuire
Journal:  Int Urol Nephrol       Date:  2012-07-21       Impact factor: 2.370

2.  Long-term outcome of transobturator tape (TOT) for treatment of stress urinary incontinence in females with neuropathic bladders.

Authors:  G S Losco; J R Burki; Y A I Omar; P J R Shah; R Hamid
Journal:  Spinal Cord       Date:  2015-04-28       Impact factor: 2.772

3.  The bone-anchor sub-urethral sling for the treatment of iatrogenic male incontinence: subjective and objective assessment after 41 months of mean follow-up.

Authors:  Claudio Giberti; Fabrizio Gallo; Maurizio Schenone; Pierluigi Cortese
Journal:  World J Urol       Date:  2007-11-03       Impact factor: 4.226

Review 4.  Treatment of postprostatectomy incontinence: is the bulbourethral sling a viable alternative to the artificial urinary sphincter?

Authors:  Steven P Petrou
Journal:  Curr Urol Rep       Date:  2002-10       Impact factor: 2.862

  4 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.