Literature DB >> 10863084

Analysis of head motion prior to and during proton beam therapy.

R W Schulte1, R A Fargo, H J Meinass, J D Slater, J M Slater.   

Abstract

PURPOSE: We report on the use of a noninvasive patient motion monitoring system to evaluate the amount of head motion prior to and during proton radiation therapy sessions. METHODS AND MATERIALS: Two optical displacement sensors, placed close to the patient's head, were used for online monitoring of the head position, with submillimeter accuracy. Motion data, including the difference between start and end position (Dx) and the maximum displacement during the recorded session (Dx-max), were acquired for pretreatment sessions to analyze alignment radiographs, and for treatment sessions. We have recorded 102 pretreatment and 99 treatment sessions in 16 patients immobilized with a thermoplastic mask, and 44 pretreatment and 56 treatment sessions in 13 patients immobilized with vacuum-assisted dental fixation. To avoid incorrect data analysis due to replicate observations, only 1 pretreatment and 1 treatment session per patient were selected at random for statistical comparison of mean or median motion parameters in different subgroups.
RESULTS: Both techniques showed similar immobilization efficiencies. The median Dx and Dx-max values were 0. 18 mm and 0.46 mm, respectively, for 16 treatment sessions with mask immobilization, and 0.22 mm and 0.50 mm, respectively, for 13 treatment sessions with dental immobilization. Motion parameters for pretreatment and treatment sessions were not statistically different.
CONCLUSION: Online verification of patient's head motion is feasible and provides valuable data for confirmation of proper treatment delivery in individual patients, as well as for the evaluation of different immobilization methods.

Entities:  

Mesh:

Substances:

Year:  2000        PMID: 10863084     DOI: 10.1016/s0360-3016(00)00551-4

Source DB:  PubMed          Journal:  Int J Radiat Oncol Biol Phys        ISSN: 0360-3016            Impact factor:   7.038


  6 in total

1.  Water equivalent thickness analysis of immobilization devices for clinical implementation in proton therapy.

Authors:  A J Wroe; A Ghebremedhin; I R Gordon; R W Schulte; J D Slater
Journal:  Technol Cancer Res Treat       Date:  2013-08-31

2.  Dosimetric robustness against setup errors in charged particle radiotherapy of skull base tumors.

Authors:  Filippo Ammazzalorso; Urszula Jelen; Rita Engenhart-Cabillic; Wolfgang Schlegel
Journal:  Radiat Oncol       Date:  2014-12-05       Impact factor: 3.481

3.  Effectiveness of base-of-skull immobilization system in a compact proton therapy setting.

Authors:  Ghazal Shafai-Erfani; Twyla Willoughby; Naren Ramakrishna; Sanford Meeks; Patrick Kelly; Omar Zeidan
Journal:  J Appl Clin Med Phys       Date:  2018-04-06       Impact factor: 2.102

4.  Utilization of optical tracking to assess efficacy of intracranial immobilization techniques in proton therapy.

Authors:  Wen C Hsi; Andries N Schreuder; Omar Zeidan
Journal:  J Appl Clin Med Phys       Date:  2015-09-08       Impact factor: 2.102

5.  Multi-scenario based robust intensity-modulated proton therapy (IMPT) plans can account for set-up errors more effectively in terms of normal tissue sparing than planning target volume (PTV) based intensity-modulated photon plans in the head and neck region.

Authors:  Martin Stuschke; Andreas Kaiser; Jehad Abu Jawad; Christoph Pöttgen; Sabine Levegrün; Jonathan Farr
Journal:  Radiat Oncol       Date:  2013-06-18       Impact factor: 3.481

6.  Migration from full-head mask to "open-face" mask for immobilization of patients with head and neck cancer.

Authors:  Guang Li; D Michael Lovelock; James Mechalakos; Shyam Rao; Cesar Della-Biancia; Howard Amols; Nancy Lee
Journal:  J Appl Clin Med Phys       Date:  2013-09-06       Impact factor: 2.102

  6 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.