Microdialysis measurement of glucose in subcutaneous adipose tissue up to three weeks in type 1 diabetic patients.

H L Lutgers1, L M Hullegie, K Hoogenberg, W J Sluiter, R P Dullaart, K J Wientjes, A J Schoonen.   

Abstract

BACKGROUND: Microdialysis of subcutaneous adipose tissue may provide an opportunity to monitor glucose continuously, when the device is connected to an extracorporal glucose sensor. We assessed whether our microdialysis probes are capable of measuring adipose tissue glucose over a prolonged period in Type 1 diabetic patients. Furthermore, the relationship between abdominal skinfold thickness and glucose recovery and the effect of spontaneous glucose excursions on its recovery were evaluated.
METHODS: Microdialysis probes were pairwise inserted subcutaneously into the abdominal fat and remained in situ for 3 weeks in eight Type 1 diabetic patients. At days 1, 3, 4, 8, 11, 16, and 18 of probe retention, glucose, as measured by microdialysis, was compared to capillary blood glucose concentrations during a 4 h period. The recovery of glucose obtained by microdialysis was expressed as a percentage of the capillary blood glucose concentration.
RESULTS: Eleven of the 16 inserted probes (69%) were evaluable during the complete study. Recovery of glucose was lower at day 1 and 3 (51+/-23% and 56+/-18%, respectively, mean+/-S.D.) compared to values found afterwards (67+/-19%, 72+/-13%, 76+/-14%, 71+/-16%, and 76+/-18%, for day 4, 8, 11, 16, and 18, respectively, for all P<0.05 vs. day 1 and 3). Skinfold thickness was inversely related to the overall 3 week glucose recovery (r=-0.76; P<0.03). Recovery was similar over a wide range of capillary blood glucose concentrations.
CONCLUSIONS: Prolonged in vivo retention of microdialysis probes improves the recovery and lowers the variability of adipose tissue-sampled glucose in Type 1 diabetic patients. These findings show that microdialysis-based glucose measurements offer an opportunity for prolonged glucose monitoring.

Entities:  

Mesh:

Substances:

Year:  2000        PMID: 10862996     DOI: 10.1016/s0300-2977(00)00022-x

Source DB:  PubMed          Journal:  Neth J Med        ISSN: 0300-2977            Impact factor:   1.422


  8 in total

1.  Analyte flux at a biomaterial-tissue interface over time: implications for sensors for type 1 and 2 diabetes mellitus.

Authors:  Neda Rajamand Ekberg; Kerstin Brismar; Jonas Malmstedt; Mari-Anne Hedblad; Ulf Adamson; Urban Ungerstedt; Natalie Wisniewski
Journal:  J Diabetes Sci Technol       Date:  2010-09-01

Review 2.  Technologies for continuous glucose monitoring: current problems and future promises.

Authors:  Santhisagar Vaddiraju; Diane J Burgess; Ioannis Tomazos; Faquir C Jain; Fotios Papadimitrakopoulos
Journal:  J Diabetes Sci Technol       Date:  2010-11-01

Review 3.  In vitro, in vivo and post explantation testing of glucose-detecting biosensors: current methods and recommendations.

Authors:  Heidi E Koschwanez; William M Reichert
Journal:  Biomaterials       Date:  2007-04-19       Impact factor: 12.479

4.  Blood-induced interference of glucose sensor function in vitro: implications for in vivo sensor function.

Authors:  Ulrike Klueh; Zenghe Liu; Tianmei Ouyang; Brian Cho; Ben Feldman; Timothy P Henning; Don Kreutzer
Journal:  J Diabetes Sci Technol       Date:  2007-11

5.  Pilot investigation on long-term subcutaneous microdialysis: proof of principle in humans.

Authors:  Franziska Simmel; Claudia Kirbs; Zeynep Erdogan; Edith Lackner; Markus Zeitlinger; Charlotte Kloft
Journal:  AAPS J       Date:  2012-10-13       Impact factor: 4.009

6.  Time lag of glucose from intravascular to interstitial compartment in humans.

Authors:  Ananda Basu; Simmi Dube; Michael Slama; Isabel Errazuriz; Jose Carlos Amezcua; Yogish C Kudva; Thomas Peyser; Rickey E Carter; Claudio Cobelli; Rita Basu
Journal:  Diabetes       Date:  2013-09-05       Impact factor: 9.461

Review 7.  Bench-to-bedside review: microdialysis in intensive care medicine.

Authors:  Stephan Klaus; Matthias Heringlake; Ludger Bahlmann
Journal:  Crit Care       Date:  2004-06-03       Impact factor: 9.097

8.  Glucose sensing in the peritoneal space offers faster kinetics than sensing in the subcutaneous space.

Authors:  Daniel R Burnett; Lauren M Huyett; Howard C Zisser; Francis J Doyle; Brett D Mensh
Journal:  Diabetes       Date:  2014-03-12       Impact factor: 9.461

  8 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.