Effect of corneal thickness on the accuracy of intraocular pressure measurement in rabbits after excimer laser photoablation.

J Stahl1, S Vold.   

Abstract

PURPOSE: To determine whether intraocular pressure (IOP) measurements using pneumotonometry and Tono-Pen tonometry are accurate after excimer laser photoablation of the central cornea in rabbits.
SETTING: Department of Ophthalmology, Scott & White Hospital, Texas A & M University College of Medicine, Temple, Texas, USA.
METHODS: Ten rabbits had excimer laser phototherapeutic keratectomy (PTK) in the right eye to thin the central corneal thickness (CCT) by approximately 20%. The left eye served as the control. The diameter of each PTK was 6.0 mm. Four weeks later, the rabbits were examined under general anesthesia. The eyes were cannulated, and IOP was maintained at 10, 20, and 30 mm Hg and measured using an IOP transducer, Tono-Pen, and pneumotonometer at each pressure level. The readings were compared.
RESULTS: One rabbit was excluded from the study because of recurrent erosions and corneal edema. There was no statistically significant difference in measurement error between the operated eye and control eye for the Tono-Pen (P = .12) or pneumotonometer (P = .35). There were significant differences in measurement error as pressure increased for the Tono-Pen (P < .01) and pneumotonometer (P < .01). The increase in measurement error with the increase in pressure was similar in the 2 eyes: Tono-Pen (P = .96) and pneumotonometer (P = .55). A paired t test showed the observed CCT was comparable (P = .79) in the pre-laser right eyes (mean 400.56 microm +/- 11.29 [SD]) and control eyes (mean 401.78 +/- 19.78 microm).
CONCLUSION: Photoablation of approximately 20% of the CCT in rabbits by excimer laser PTK did not significantly alter the accuracy of IOP measurements by the Tono-Pen or pneumotonometer.

Entities:  

Mesh:

Year:  2000        PMID: 10831906     DOI: 10.1016/s0886-3350(99)00463-0

Source DB:  PubMed          Journal:  J Cataract Refract Surg        ISSN: 0886-3350            Impact factor:   3.351


  2 in total

1.  The effects of the modulus of the lens material on intraocular pressure measurement through soft contact lenses.

Authors:  S Boyraz; I Güngör
Journal:  Ir J Med Sci       Date:  2012-12-04       Impact factor: 1.568

2.  A comparison between Goldmann applanation tonometry and dynamic contour tonometry after photorefractive keratectomy.

Authors:  Afshin Lotfi Sadigh; Rohollah F Fouladi; Hassan Hashemi; Amir Houshang Beheshtnejad
Journal:  Graefes Arch Clin Exp Ophthalmol       Date:  2012-09-02       Impact factor: 3.117

  2 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.