Uniport soft-tip, open-ended catheters versus multiport firm-tipped close-ended catheters for epidural labor analgesia: a quality assurance study.

F Jaime1, G L Mandell, M C Vallejo, S Ramanathan.   

Abstract

STUDY
OBJECTIVE: To compare a multiport, firm-tipped, close-ended, epidural catheter (Portex catheter) with a uniport, open-ended, soft-tipped, wire-reinforced catheter (Arrow catheter) in obstetric patients. STUDY
DESIGN: Prospective data collection for intradepartmental quality assurance.
SETTING: Obstetric unit in a tertiary care maternity hospital. PATIENTS: 2612 patients requesting labor analgesia.
INTERVENTIONS: The Arrow catheter was used in 1,352 women and the Portex catheter in 1,260 women.
MEASUREMENTS AND MAIN RESULTS: The incidence of unsatisfactory block were 3.3% and 4.4% with the Arrow and Portex catheters, respectively (p = 0.2). The catheter perforated the dura matter in 0.4% of cases with both catheters. The incidence of epidural venipuncture was 1.1% with the Arrow catheter and 5.7% with the Portex catheter (p = 0.0001). Paresthesias occurred in 6% of cases with the Arrow catheter and 11. 2 % of cases with the Portex catheter (p = 0.0001). Epidural catheter reinsertion was required in fewer patients in the Arrow group than in the Portex group (4.8% vs. 7.1%; p = 0.01).
CONCLUSIONS: In obstetric patients, the softer uniport Arrow catheter produces paresthesias and venipunctures less frequently than the firm multiport Portex catheter.

Entities:  

Mesh:

Substances:

Year:  2000        PMID: 10818320     DOI: 10.1016/s0952-8180(00)00116-1

Source DB:  PubMed          Journal:  J Clin Anesth        ISSN: 0952-8180            Impact factor:   9.452


  4 in total

1.  Why epidurals do not always work.

Authors:  Katherine Arendt; Scott Segal
Journal:  Rev Obstet Gynecol       Date:  2008

2.  Prompt Detection of Intravascular Migration of Cervical Epidural Catheter in Superior Vena Cava Avoids Impending Disaster!

Authors:  Sandeep Diwan; Abhijit Nair
Journal:  Turk J Anaesthesiol Reanim       Date:  2021-04-01

3.  Paths of thoracic epidural catheters in children undergoing the Nuss procedure for pectus excavatum repair.

Authors:  Kanna Nakamura; Ayanori Sugita; Shuichi Sekiya; Akira Kitamura; Hiromasa Mitsuhata; Keisuke Yamaguchi; Masakazu Hayashida
Journal:  J Anesth       Date:  2022-03-04       Impact factor: 2.931

4.  Lateral deviation of four types of epidural catheters from the lumbar epidural space into the intervertebral foramen.

Authors:  Tetsuya Uchino; Masahiro Miura; Yoshimasa Oyama; Shigekiyo Matsumoto; Chihiro Shingu; Takaaki Kitano
Journal:  J Anesth       Date:  2016-04-30       Impact factor: 2.078

  4 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.