Reliability of intraocular lens power calculation after cataract surgery in patients with relative anterior microphthalmos.

T U Bartke1, G U Auffarth, J C Uhl, H E Völcker.   

Abstract

BACKGROUND: Patients with relative anterior microphthalmus (RAM) are characterized by special anatomical features (corneal diameters <11 mm, anterior chamber depth <2 mm and normal axial length) that sometimes make intraocular lens (IOL) power calculation difficult. PATIENTS AND METHODS: Seventy-five patients aged 75.6+/-10.3 years with RAM were evaluated for preoperative target refraction and postoperative refraction after cataract surgery. We used biometric formulas as modified by Haigis for IOL power calculation.
RESULTS: The average IOL power implanted was 25.2+/-2.8 dptr (one-piece all-PMMA IOL; range 19-31 dptr). The preoperatively calculated target refraction was -0.71+/-1.43 dptr (range -4.3 to +2.2 dptr). The average postoperative spherical equivalent was -0.41+/-1.50 dptr (range -5.5 to +6.0 dptr). The mean difference between target and end refraction was -0.30+/-1.54 dptr Fifty-seven percent of cases did not differ by more than 1 dptr from target refraction, 81.3% by not more than 2 dptr, and 94.7% by not more than 3 dptr; only 5.3% of cases deviated by more than 3 dptr. There was no correlation of the amount of deviation between target and postoperative refraction with axial length, corneal diameter and anterior chamber depth. There was, however, a significant correlation between target refraction and amount of deviation of spherical equivalent. The largest differences were found with target refractions greater than +2 dptr.
CONCLUSIONS: In spite of the special anatomical conditions in patients with RAM the biometric formulas as modified by Haigis produced reliable IOL power calculations. The best accuracy was achieved when aiming at a target refraction in the range of +/-2 dptr.

Entities:  

Mesh:

Substances:

Year:  2000        PMID: 10766282     DOI: 10.1007/pl00007881

Source DB:  PubMed          Journal:  Graefes Arch Clin Exp Ophthalmol        ISSN: 0721-832X            Impact factor:   3.117


  2 in total

Review 1.  [IOL calculation for high ametropia].

Authors:  W Haigis
Journal:  Ophthalmologe       Date:  2008-11       Impact factor: 1.059

2.  Posterior chamber intraocular lens implantation in pediatric cataract with microcornea and/or microphthalmos.

Authors:  Young Suk Yu; Seong-Joon Kim; Ho Kyoung Choung
Journal:  Korean J Ophthalmol       Date:  2006-09
  2 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.