Literature DB >> 10760439

Assessment of the uncertainties in dose delivery of a commercial system for linac-based stereotactic radiosurgery.

D Verellen1, N Linthout, A Bel, G Soete, D van den Berge, J D' Haens, G Storme.   

Abstract

PURPOSE: Linac-based stereotactic radiosurgery (SRS) was introduced in our department in 1992, and since then, more than 200 patients have been treated with this method. An in-house-developed algorithm for target localization and dose calculation has recently been replaced with a commercially available system. In this study, both systems have been compared, and positional accuracy, as well as dose calculation, have been verified experimentally. METHODS AND MATERIALS: The in-house-developed software for target localization and dose calculation is an extension to George Sherouse's GRATIS(R) software for radiotherapy treatment planning, and has been replaced by a commercial (BrainSCAN version 3.1; BrainLAB, Germany) treatment planning system (TPS) for SRS. The positional accuracy for the entire SRS procedure (from image acquisition to treatment) has been investigated by treatment of simulated targets in the form of 0.2-cm lead beads inserted into an anthropomorphic phantom. Both dose calculation algorithms have been verified against manual calculations (based on basic beam data and CT data from phantom and patients), and measurements with the anthropomorphic phantom applying ionization chamber, thermoluminescent detectors, and radiographic film. This analysis has been performed on a variety of experimental situations, starting with static beams and simple one-arc treatments, to more complex and clinical relevant applications. Finally, 11 patients have been evaluated with both TPS in parallel for comparison and continuity of clinical experience.
RESULTS: Phantom studies evaluating the entire SRS procedure have shown that a target, localized by CT, can be irradiated with a positional accuracy of 0.08 cm in any direction with 95% confidence. Neglecting the influence of dose perturbation when the beam passes through bone tissue or air cavities, the calculated dose values obtained from both TPSs agreed within 1% (SD 1%) for phantom and patient studies. The application of a one-dimensional path length correction for tissue heterogeneity influences the treatment prescription 4% on average (SD 1%), which is in compliance with theoretical predictions. The phantom measurements confirmed the predicted dose at isocenter within uncertainty for the different treatment schedules in this study.
CONCLUSION: The full SRS procedure applied to an anthropomorphic phantom has been used as a comprehensive method to assess the uncertainties involved in dose delivery and target positioning. The results obtained with both TPSs are in agreement with AAPM Report 54, TG 42 and clinical continuity is assured. However, the use of a one-dimensional path length correction will result in an increase of 4% in dose prescription, which is slightly more than that predicted in the literature.

Entities:  

Mesh:

Year:  1999        PMID: 10760439     DOI: 10.1016/s0360-3016(99)00020-6

Source DB:  PubMed          Journal:  Int J Radiat Oncol Biol Phys        ISSN: 0360-3016            Impact factor:   7.038


  8 in total

1.  Dosimetric study on learning-based cone-beam CT correction in adaptive radiation therapy.

Authors:  Tonghe Wang; Yang Lei; Nivedh Manohar; Sibo Tian; Ashesh B Jani; Hui-Kuo Shu; Kristin Higgins; Anees Dhabaan; Pretesh Patel; Xiangyang Tang; Tian Liu; Walter J Curran; Xiaofeng Yang
Journal:  Med Dosim       Date:  2019-04-01       Impact factor: 1.482

2.  In patient dose reconstruction using a cine acquisition for dynamic arc radiation therapy.

Authors:  Angelo Piermattei; Andrea Fidanzio; Luigi Azario; Francesca Greco; Alessandra Mameli; Savino Cilla; Luca Grimaldi; Guido D'Onofrio; Boris Giuseppe Augelli; Gerardina Stimato; Diego Gaudino; Sara Ramella; Rolando D'Angelillo; Francesco Cellini; Lucio Trodella
Journal:  Med Biol Eng Comput       Date:  2009-02-17       Impact factor: 2.602

3.  A preplanning method for stereotactic radiosurgery to improve treatment workflow.

Authors:  Kang-Hyun Ahn; Naim Ozturk; Brett Smith; Konstantin V Slavin; Matthew Koshy; Bulent Aydogan
Journal:  J Appl Clin Med Phys       Date:  2016-05-08       Impact factor: 2.102

4.  An immobilization and localization technique for SRT and IMRT of intracranial tumors.

Authors:  Leonid B Leybovich; Anil Sethi; Nesrin Dogan; Edward Melian; Mathew Krasin; Bahman Emami
Journal:  J Appl Clin Med Phys       Date:  2002       Impact factor: 2.102

5.  Six degrees of freedom CBCT-based positioning for intracranial targets treated with frameless stereotactic radiosurgery.

Authors:  Anees Dhabaan; Eduard Schreibmann; Arsalan Siddiqi; Eric Elder; Tim Fox; Tomi Ogunleye; Natia Esiashvili; Walter Curran; Ian Crocker; Hui-Kuo Shu
Journal:  J Appl Clin Med Phys       Date:  2012-11-08       Impact factor: 2.102

Review 6.  Isocenter verification for linac-based stereotactic radiation therapy: review of principles and techniques.

Authors:  Pejman Rowshanfarzad; Mahsheed Sabet; Daryl J O'Connor; Peter B Greer
Journal:  J Appl Clin Med Phys       Date:  2011-11-15       Impact factor: 2.102

7.  Feasibility of using the Vero SBRT system for intracranial SRS.

Authors:  Manuela Burghelea; Dirk Verellen; Thierry Gevaert; Tom Depuydt; Kenneth Poels; Viorica Simon; Mark De Ridder
Journal:  J Appl Clin Med Phys       Date:  2014-01-06       Impact factor: 2.102

8.  End to end comparison of surface-guided imaging versus stereoscopic X-rays for the SRS treatment of multiple metastases with a single isocenter using 3D anthropomorphic gel phantoms.

Authors:  Victoria Bry; Daniel Saenz; Evangelos Pappas; Georgios Kalaitzakis; Nikos Papanikolaou; Karl Rasmussen
Journal:  J Appl Clin Med Phys       Date:  2022-03-24       Impact factor: 2.243