| Literature DB >> 10743781 |
R Grieve1, P A Parker, B Hudgins, K Englehart.
Abstract
Noninvasive measurements of somatosensory evoked potentials have both clinical and research applications. The electrical artifact which results from the stimulus is an interference which can distort the evoked signal, and introduce errors in response onset timing estimation. Given that this interference is synchronous with the evoked signal, it cannot be reduced by the conventional technique of ensemble averaging. The technique of adaptive noise cancelling has potential in this regard however, and has been used effectively in other similar problems. An adaptive noise cancelling filter which uses a neural network as the adaptive element is investigated in this application. The filter is implemented and performance determined in the cancelling of artifact for in vivo measurements on the median nerve. A technique of segmented neural network training is proposed in which the network is trained on that segment of the record time window which does not contain the evoked signal. The neural network is found to generalize well from this training to include the segment of the window containing the evoked signal. Both quantitative and qualitative measures show that significant stimulus artifact reduction is achieved.Mesh:
Year: 2000 PMID: 10743781 DOI: 10.1109/10.827307
Source DB: PubMed Journal: IEEE Trans Biomed Eng ISSN: 0018-9294 Impact factor: 4.538