Literature DB >> 10735407

Downstream turbulence and high intensity transient signals (HITS) following aortic valve replacement with Medtronic Hall or St. Jude Medical valve substitutes.

P Kleine1, M Perthel, J M Hasenkam, H Nygaard, S B Hansen, J Laas.   

Abstract

OBJECTIVE: High intensity transient signals (HITS) representing microembolization to the brain have been found to contribute to cognitive impairment and psychoneurological dysfunction in patients carrying a mechanical aortic valve. It is unknown, whether HITS represent gaseous or solid emboli. This animal study evaluates the impact of valve orientation on HITS for two different mechanical valves with both valves implanted in their best and worst orientation, which has been defined in previous studies with respect to downstream turbulence.
METHODS: In four pigs a rotation device carrying either a Medtronic Hall (MH) or St. Jude Medical (SJM) valve size 23 mm was implanted. The device allowed rotation of the implanted valves without reopening of the aorta. Approximately 30 min after weaning from extracorporeal circulation, a Doppler probe was placed on both common carotid arteries. In different orientations of the implanted valves (best and worst position), HITS were detected by the Doppler probe and recorded for ten min by a transcranial Doppler sonography device (Medilab Inc., Estenfeld, Germany).
RESULTS: HITS showed significant change with rotation for both valve designs. With the major orifice of the MH oriented towards the non-coronary leaflet (optimum position) very low HITS-counts (0.8-1.7/min) were observed. In the worst orientation HITS rose to 43-66/min. For the SJM the HITS count in the optimum position was 23.4-24/min and in the worst orientation 38-48/min.
CONCLUSIONS: Valve orientation has an important impact on microembolization to the brain. In the optimum orientation (large orifice facing the non-coronary leaflet) the Medtronic Hall valve showed negligible incidence of HITS. The St. Jude Medical bileaflet valve showed less variation but demonstrated significant HITS counts at any orientation. As the MH in the worst position shows significantly higher turbulent stresses than the SJM but no higher incidence of HITS, a strong correlation between turbulence and HITS was not demonstrated by this study.

Entities:  

Mesh:

Substances:

Year:  2000        PMID: 10735407     DOI: 10.1016/s1010-7940(99)00351-6

Source DB:  PubMed          Journal:  Eur J Cardiothorac Surg        ISSN: 1010-7940            Impact factor:   4.191


  6 in total

1.  The effect of implantation orientation of a bileaflet mechanical heart valve on kinematics and hemodynamics in an anatomic aorta.

Authors:  Iman Borazjani; Fotis Sotiropoulos
Journal:  J Biomech Eng       Date:  2010-11       Impact factor: 2.097

2.  Embolic and bleeding events after modified Bentall procedure in selected patients.

Authors:  N C Radu; E W M Kirsch; M-L Hillion; F Lagneau; L Drouet; D Loisance
Journal:  Heart       Date:  2006-06-27       Impact factor: 5.994

3.  [Turbulence and high intensity transient signals (HITS) as a parameter for optimum orientation of mechanical heart valves].

Authors:  M Perthel; J M Hasenkam; H Nygaard; W Kupper; J Laas
Journal:  Z Kardiol       Date:  2001-12

4.  Effects of Bileaflet Mechanical Mitral Valve Rotational Orientation on Left Ventricular Flow Conditions.

Authors:  John C Westerdale; Ronald Adrian; Kyle Squires; Hari Chaliki; Marek Belohlavek
Journal:  Open Cardiovasc Med J       Date:  2015-06-26

5.  Relationship between exercise-induced heart rate increase and the formation of microbubbles and high-intensity transient signals in mechanical heart valve implanted patients.

Authors:  Ayşegül Sünbül; Ahmet Kırbaş; Nursen Tanrıkulu; Cihan Sengül; Bahadır Dağdeviren; Omer Işık
Journal:  Arch Med Sci       Date:  2013-04-30       Impact factor: 3.318

6.  Effects of pannus formation on the flow around a bileaflet mechanical heart valve.

Authors:  Woojin Kim; Haecheon Choi; Jihoon Kweon; Dong Hyun Yang; Young-Hak Kim
Journal:  PLoS One       Date:  2020-06-12       Impact factor: 3.240

  6 in total