Literature DB >> 10714779

Biomechanical analysis of supracondylar femoral fractures fixed with modern retrograde intramedullary nails.

N Hora1, D C Markel, A Haynes, M J Grimm.   

Abstract

OBJECTIVES: Several new retrograde supracondylar intramedullary nails have been developed to specifically address fractures of the distal femur. The nails appear clinically effective, but there are few biomechanical data documenting the stability of the fixation or the mechanical stiffness of the different designs. The goal of this study was to assess the torsional and bending stiffness of four designs of intramedullary nails developed for this application.
METHODS: Four nail designs were tested in torsion and bending to determine system stiffness: Ace supracondylar, Richards "five hole" and "multi-hole" supracondylar, and Biomet retrograde. The nails were inserted into cadaveric femurs in which a one-centimeter distraction osteotomy had been created seven centimeters proximal to the condyles. The constructs were then tested on an Instron biaxial testing system.
RESULTS: There were no statistically significant differences in bending stiffness among the groups of nails (range 0.79 to 1.18 newtons/meter; p > 0.1). However, the Ace nails (1.10 newtonmeters/degree) did exhibit a statistically lower torsional stiffness compared with the other nails (2.20 to 2.21 newton-meters/ degree; p < 0. 1). No differences were noted as a function of the number of locking holes.
CONCLUSIONS: The bending stiffness of four currently available designs of retrograde intramedullary nails does not appear to be dependent on design variations. The torsional stiffness did vary among the four designs, but this was not determined by the number of fixation holes provided. It appears that a well-placed retrograde supracondylar nail of modern design should have sufficient stiffness to support the femur and provide stability during fracture healing.

Entities:  

Mesh:

Year:  1999        PMID: 10714779     DOI: 10.1097/00005131-199911000-00004

Source DB:  PubMed          Journal:  J Orthop Trauma        ISSN: 0890-5339            Impact factor:   2.512


  4 in total

1.  [Femoral nail osteosynthesis. Mechanical factors influencing the femoral antetorsion].

Authors:  M Citak; D Kendoff; M Citak; M J Gardner; M Oszwald; C Krettek; T Hüfner
Journal:  Unfallchirurg       Date:  2008-04       Impact factor: 1.000

2.  [Post-traumatic torsional differences and functional tests following antegrade or retrograde intramedullary nailing of the distal femoral diaphysis].

Authors:  D G Maier; R Reisig; P Keppler; L Kinzl; F Gebhard
Journal:  Unfallchirurg       Date:  2005-02       Impact factor: 1.000

3.  Prospective comparative study of two methods for fixation after distal femur corrective osteotomy for valgus deformity; retrograde intramedullary nailing versus less invasive stabilization system plating.

Authors:  Çağrı Özcan; Sami Sökücü; Kubilay Beng; Engin Çetinkaya; Bilal Demir; Yavuz Selim Kabukçuoğlu
Journal:  Int Orthop       Date:  2016-04-14       Impact factor: 3.075

4.  Temperature influence on DXA measurements: bone mineral density acquisition in frozen and thawed human femora.

Authors:  Dirk Wähnert; Konrad L Hoffmeier; Gabriele Lehmann; Rosemarie Fröber; Gunther O Hofmann; Thomas Mückley
Journal:  BMC Musculoskelet Disord       Date:  2009-02-24       Impact factor: 2.362

  4 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.