Are routine arrhythmia inductions necessary in patients with pectoral implantable cardioverter defibrillators?

M Glikson1, D Luria, P A Friedman, J M Trusty, M Benderly, S C Hammill, M S Stanton.   

Abstract

INTRODUCTION: The value of ventricular arrhythmia inductions as part of routine implantable cardioverter defibrillator (ICD) follow-up in new-generation pectoral ICDs is unknown. METHODS AND
RESULTS: We performed a retrospective analysis of a prospectively collected database analyzing data from 153 patients with pectoral ICDs who had routine arrhythmia inductions at predismissal, and 3 months and 1 year after implantation. Routine predismissal ventricular fibrillation (VF) induction yielded important findings in 8.8% of patients, all in patients with implantation defibrillation threshold (DFT) > or = 15 J or with concomitant pacemaker systems. At 3 months and 1 year, routine VF induction yielded important findings in 5.9% and 3.8% of tested patients, respectively, all in patients who had high DFT on prior testing. Ventricular tachycardia (VT) induction at predismissal, and 3 months and 1 year after implantation resulted in programming change in 37.4%, 28.1%, and 13.8% of tested patients, almost all in patients with inducible VT on baseline electrophysiologic study and clinical episodes since implantation.
CONCLUSION: Although helpful in identifying potentially important ICD malfunctions, routine arrhythmia inductions during the first year after ICD implantation may not be necessary in all cases. VF inductions have a low yield in patients with previously low DFTs who lack concomitant pacemakers. VT inductions have a low yield in patients without baseline inducible VT and in the absence of clinical events. Definite recommendations regarding patient selection must await larger prospective studies as well as consensus in the medical community about what comprises an acceptable risk justifying avoidance of the costs and inconveniences of routine arrhythmia inductions.

Entities:  

Mesh:

Year:  2000        PMID: 10709706     DOI: 10.1111/j.1540-8167.2000.tb00311.x

Source DB:  PubMed          Journal:  J Cardiovasc Electrophysiol        ISSN: 1045-3873


  7 in total

1.  A protocol for patients with cardiovascular implantable devices undergoing magnetic resonance imaging (MRI): should defibrillation threshold testing be performed post-(MRI).

Authors:  Peter Thomas Burke; Hamid Ghanbari; Patrick B Alexander; Michael K Shaw; Marcos Daccarett; Christian Machado
Journal:  J Interv Card Electrophysiol       Date:  2010-01-29       Impact factor: 1.900

2.  Safety of Magnetic Resonance Imaging in Patients with Cardiac Devices.

Authors:  Saman Nazarian; Rozann Hansford; Amir A Rahsepar; Valeria Weltin; Diana McVeigh; Esra Gucuk Ipek; Alan Kwan; Ronald D Berger; Hugh Calkins; Albert C Lardo; Michael A Kraut; Ihab R Kamel; Stefan L Zimmerman; Henry R Halperin
Journal:  N Engl J Med       Date:  2017-12-28       Impact factor: 91.245

3.  A prospective evaluation of a protocol for magnetic resonance imaging of patients with implanted cardiac devices.

Authors:  Saman Nazarian; Rozann Hansford; Ariel Roguin; Dorith Goldsher; Menekhem M Zviman; Albert C Lardo; Brian S Caffo; Kevin D Frick; Michael A Kraut; Ihab R Kamel; Hugh Calkins; Ronald D Berger; David A Bluemke; Henry R Halperin
Journal:  Ann Intern Med       Date:  2011-10-04       Impact factor: 25.391

4.  Clinical utility and safety of a protocol for noncardiac and cardiac magnetic resonance imaging of patients with permanent pacemakers and implantable-cardioverter defibrillators at 1.5 tesla.

Authors:  Saman Nazarian; Ariel Roguin; Menekhem M Zviman; Albert C Lardo; Timm L Dickfeld; Hugh Calkins; Robert G Weiss; Ronald D Berger; David A Bluemke; Henry R Halperin
Journal:  Circulation       Date:  2006-09-11       Impact factor: 29.690

5.  Indications for predismissal testing with arrhythmia-induction in patients receiving an implantable cardioverter defibrillator.

Authors:  G Christ; R Becker; F Voss; K Kelemen; J Senges-Becker; M Hauck; W Schoels; I Bald; H A Katus; A Bauer
Journal:  Clin Res Cardiol       Date:  2007-06-27       Impact factor: 5.460

6.  Successful implantation of cardiac defibrillators without induction of ventricular fibrillation using upper limit of vulnerability testing.

Authors:  Ulrika Birgersdotter Green; Ashok Garg; Fawzia Al-Kandari; GilAnthony Ungab; Linda Tone; Gregory K Feld
Journal:  J Interv Card Electrophysiol       Date:  2003-02       Impact factor: 1.900

7.  Defibrillation testing of the implantable cardioverter defibrillator: when, how, and by whom?

Authors:  Luis A Pires
Journal:  Indian Pacing Electrophysiol J       Date:  2007-08-01
  7 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.