A new technique for cranioplasty with L-shaped titanium plates and combination ceramic implants composed of hydroxyapatite and tricalcium phosphate (Ceratite).

H Miyake1, T Ohta, H Tanaka.   

Abstract

OBJECTIVE: The use of hydroxyapatite-based ceramics for cranioplasties has recently increased in Japan, because of the good cosmetic outcomes, biocompatibility, strength, osteoconductive properties, and lack of risk of disease transmission associated with these materials. However, miniplate fixation has not been possible for ceramic implants. We describe a new technique for miniplate fixation of ceramic implants.
METHODS: Combination ceramic implants composed of hydroxyapatite and tricalcium phosphate (Ceratite; NGK Spark Plug Co., Aichi, Japan) were used for cranioplasties. A slot and a pair of holes were cut in each Ceratite implant, for use as a fixation unit. We have also developed a new L-shaped titanium plate (HOMS Engineering Inc., Nagano, Japan) that fits into the fixation unit. We first insert an L-shaped titanium plate through the slot from the back surface of the Ceratite implant. We then bend the plate outward at the front surface of the Ceratite implant and fix it to the cranium of the patient with titanium screws. The Ceratite implant is usually firmly fixed to the cranium of the patient with three L-shaped titanium plates.
RESULTS: Using L-shaped titanium plates and Ceratite implants, we successfully performed cranioplasties for seven patients with cranial defects resulting from external decompression craniotomies. The Ceratite implant exactly fit the bone window for each patient. Surgical maneuvers were simple and easy for all patients, permitting shorter operating times. All Ceratite implants were firmly fixed, and no postoperative infections have occurred.
CONCLUSION: Our new technique for cranioplasty is simple and allows rigid fixation of Ceratite implants.

Entities:  

Mesh:

Substances:

Year:  2000        PMID: 10690731     DOI: 10.1097/00006123-200002000-00028

Source DB:  PubMed          Journal:  Neurosurgery        ISSN: 0148-396X            Impact factor:   4.654


  4 in total

1.  A review of protein adsorption on bioceramics.

Authors:  Kefeng Wang; Changchun Zhou; Youliang Hong; Xingdong Zhang
Journal:  Interface Focus       Date:  2012-03-22       Impact factor: 3.906

2.  Immediate single-stage reconstruction of complex frontofaciobasal injuries: part I.

Authors:  Akram Mohamed Awadalla; Hichem Ezzeddine; Naglaaa Fawzy; Mohammad Al Saeed; Mohammad R Ahmad
Journal:  J Neurol Surg B Skull Base       Date:  2014-10-07

Review 3.  Calvarial Reconstruction.

Authors:  Arvind Badhey; Sameep Kadakia; Moustafa Mourad; Jared Inman; Yadranko Ducic
Journal:  Semin Plast Surg       Date:  2017-10-25       Impact factor: 2.314

4.  Reconstruction of a large complex skull defect in a child: a case report and literature review.

Authors:  Ricardo Santos de Oliveira; Rodrigo Brigato; João Flávio Gurjão Madureira; Antonio Augusto Velasco Cruz; Francisco Veríssimo de Mello Filho; Nivaldo Alonso; Helio Rubens Machado
Journal:  Childs Nerv Syst       Date:  2007-07-14       Impact factor: 1.475

  4 in total