Abstract

The presence of heterophilic antibodies in the serum of a small subpopulation of individuals continues to cause false results for modern-day immunoassays. In order to determine the frequency of heterophilic antibody (HA)-related false positives within our population of positive cardiac troponin I (cTnI) patients, we assayed 200 samples using the original in-house cTnI assay (Abbott AxSYM) and the Bayer ACS:180 cTnI, which we had previously observed to be more effective at blocking HA interference. Four samples were identified as false positives based on discordant results between the two assays, as well as the correction of the false positives by treatment of the samples with heterophilic antibody blocking reagent (HBR). An 'enhanced' version of the AxSYM cTnI reagent was designed to greatly reduce or eliminate HA interference, and has now replaced the original reagents. The present study shows that the enhanced reagent significantly reduced or eliminated much of the HA interference. Comparative studies between the enhanced cTnI reagent and the original Abbott AxSYM cTnI reagent showed excellent correlation and equivalent diagnostic concordance, when HA samples were excluded from the analysis.

Entities:  

Mesh:

Substances:

Year:  2000        PMID: 10686273     DOI: 10.1016/s0009-8981(99)00260-0

Source DB:  PubMed          Journal:  Clin Chim Acta        ISSN: 0009-8981            Impact factor:   3.786


  6 in total

1.  A non-ischaemic cause of elevated troponin.

Authors:  Guy Lefort; Claude D'Antonio; Terrell Caffery
Journal:  BMJ Case Rep       Date:  2014-04-04

2.  Interferences in immunoassay.

Authors:  Jill Tate; Greg Ward
Journal:  Clin Biochem Rev       Date:  2004-05

3.  Nanoparticle-based bio-barcode assay redefines "undetectable" PSA and biochemical recurrence after radical prostatectomy.

Authors:  C Shad Thaxton; Robert Elghanian; Audrey D Thomas; Savka I Stoeva; Jae-Seung Lee; Norm D Smith; Anthony J Schaeffer; Helmut Klocker; Wolfgang Horninger; Georg Bartsch; Chad A Mirkin
Journal:  Proc Natl Acad Sci U S A       Date:  2009-10-19       Impact factor: 11.205

Review 4.  On the Effect of Heterophilic Antibodies on Serum Levels of Cardiac Troponins: A Brief Descriptive Review.

Authors:  Aleksey Michailovich Chaulin
Journal:  Life (Basel)       Date:  2022-07-24

5.  Beckman Access versus the Bayer ACS:180 and the Abbott AxSYM cardiac Troponin-I real-time immunoassays: an observational prospective study.

Authors:  Robert H Baevsky; Rajesh K Kapur; Howard A Smithline
Journal:  BMC Emerg Med       Date:  2004-07-13

6.  Recovery of spiked troponin I in four routine assays.

Authors:  Tze Ping Loh; Xiong Chang Lim; Karize Kieu; Haressh Sajiir; Siew Fong Neo; Wan Ling Cheng; Sunil Kumar Sethi
Journal:  Biochem Med (Zagreb)       Date:  2016       Impact factor: 2.313

  6 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.