| Literature DB >> 27346968 |
Tze Ping Loh1, Xiong Chang Lim1, Karize Kieu1, Haressh Sajiir1, Siew Fong Neo1, Wan Ling Cheng1, Sunil Kumar Sethi1.
Abstract
INTRODUCTION: This study aimed to examine the recovery of spiked human cardiac troponin I (cTnI) results measured by four routine assays, and investigate possible interference from microclots.Entities:
Keywords: cardiovascular diseases; clinical chemistry; immunoassay; troponin
Mesh:
Substances:
Year: 2016 PMID: 27346968 PMCID: PMC4910266 DOI: 10.11613/BM.2016.025
Source DB: PubMed Journal: Biochem Med (Zagreb) ISSN: 1330-0962 Impact factor: 2.313
Technical specifications reported by the manufacturers’ product inserts of the assays used
| Vitros TnI ES | 12 | 12 | 40 | 27 | 4.3% at 70 ng/L | 23.1 |
| Beckman Coulter Accu TnI+3 | 8 | 40 | 40 | 40 | 8.0% at 50 ng/L | 40.6 |
| Siemens cTnI-Ultra | 6 | 30 | 40 | 30 | 5.3% at 80 ng/L | 27.7 |
| Roche TnI STAT | 160 | 300 | 160 | 234 | 4.8% at 323 ng/L | 25.4 |
| CV - coefficient of variation. | ||||||
Pre-spiking samples above the limit of quantification of the three assays investigated and their concentrations.
| Beckman Coulter AccuTnI+3 | 2 | 60 | 25 | 42–77 | 2 | 42–77 |
| Siemens cTnI-Ultra | 13 | 39 | 11 | 33–68 | 2 | 51–68 |
| Roche TnI STAT | 3 | 410 | 80 | 330–480 | 3 | 330–480 |
Figure 1Box and whisker plots showing the post-spiking cardiac troponin I result in the four assays. The y-axis is in logarithmic scale and the grey horizontal bar represents the upper reference limit of the respective assays. Values below the figure summarize the median and range of concentration of troponin I in the post-spiking samples, the median recovery (%) and results (%) above the upper reference limit.
Samples exceeding the critical difference (at a predefined probability of 0.0005).
| Vitros TnI ES | 23.1 | 95 (21) | 234 (51) | 128 (28) |
| Beckman Coulter AccuTnI+3 | 40.6 | 23 (5) | 426 (93) | 8 (2) |
| Siemens cTnI-Ultra | 27.7 | 115 (25) | 283 (62) | 59 (13) |
| Roche TnI STAT | 25.4 | 37 (8) | 375 (82) | 45 (10) |
Pre- and post-re-centrifugation cardiac troponin I results.
| Mean | 35 | 36 | - 3 |
| Median | 32 | 33 | - 3 |
| Minimum | 12 | 13 | - 28 |
| Maximum | 76 | 77 | 13 |
| Percent difference (%) was calculated as [ (pre-re-centrifugation result - post-centrifugation result / pre-re-centrifugation result) × 100]. | |||