Investigations into the use of an ultrasonic chisel to cut bone. Part 1: Forces applied by clinicians.

B S Khambay1, A D Walmsley.   

Abstract

OBJECTIVES: To measure in vitro the direction and force of applied loads applied by clinicians when using both a conventional slow surgical handpiece (CH) and an ultrasonic chisel (USC) for cutting bone. STUDY
DESIGN: Five clinicians were asked to cut bovine bone using either an USC or a CH. The bone was placed on a force measurement system that could measure both longitudinal and downward loads. The rate of cut was calculated over a fixed time-period and the depth of cut measured using a penetratometer.
RESULTS: The magnitude of the longitudinal forces generated varied between 1.48 and 3.22 N (USC) and 0.04 and 4.56 N (CH). The CH had a pulling force directed towards the operator. Both instruments produced a similar range of downward forces although there was intra- and inter-operator variability. The rate of cut varied in a similar manner, however, the CH produced a significantly greater depth of cut (p < 0.05).
CONCLUSIONS: The force measurement system demonstrated differences in the way clinicians used the USC and CH instruments to cut bone. Of the two cutting methods investigated, the rotary bur is more efficient than the ultrasonic chisel. An ultrasonic chisel does cut bone in a different manner from a conventional bur and clinicians may require training before using it clinically.

Entities:  

Mesh:

Year:  2000        PMID: 10666959     DOI: 10.1016/s0300-5712(99)00043-3

Source DB:  PubMed          Journal:  J Dent        ISSN: 0300-5712            Impact factor:   4.379


  4 in total

1.  Piezoelectric osteotomy: a new technique for bone surgery-advantages in craniofacial surgery.

Authors:  Arnaud Gleizal; Jean-Christophe Bera; Bernard Lavandier; Jean-Luc Beziat
Journal:  Childs Nerv Syst       Date:  2007-03-14       Impact factor: 1.475

2.  Characterisation of bone following ultrasonic cutting.

Authors:  Sunil Claire; Simon C Lea; A Damien Walmsley
Journal:  Clin Oral Investig       Date:  2012-05-26       Impact factor: 3.573

3.  Osteoblast viability and detachment following exposure to ultrasound in vitro.

Authors:  H Sura; R M Shelton; A D Walmsley
Journal:  J Mater Sci Mater Med       Date:  2001 Oct-Dec       Impact factor: 3.896

4.  Understanding nonlinear vibration behaviours in high-power ultrasonic surgical devices.

Authors:  Andrew Mathieson; Andrea Cardoni; Niccolò Cerisola; Margaret Lucas
Journal:  Proc Math Phys Eng Sci       Date:  2015-04-08       Impact factor: 2.704

  4 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.