Evaluation of frequency and type of errors detected by a computerized record and verify system during radiation treatment.

N Barthelemy-Brichant1, J Sabatier, W Dewé, A Albert, J M Deneufbourg.   

Abstract

BACKGROUND: Computerized record and verify systems (RVS) have been introduced to improve the precision of radiation treatment delivery. These systems prevent the delivery of ionizing radiations when the settings of the treatment machine do not match the intended parameters within some maximal authorized deviation.
PURPOSE: To assess the potential alteration of the frequency of errors associated with the use of RVS during radiation treatment delivery.
MATERIALS AND METHODS: The software of the RVS was altered in order to record the settings actually used for radiation treatment delivery whereas the verification function was suppressed. At the end of the study period, the settings used during daily administration of radiation treatment were compared to the parameters recorded in the RVS using the computer. They were also compared with the planned ones written in the patient treatment chart.
RESULTS: Out of the 147,476 parameters examined during the study period, 678 (0.46%) were set erroneously. At least one error occurred in 628 (3.22%) of the 19,512 treated fields. An erroneous parameter was introduced in the RVS memory in 22 (1.17%) of the 1885 fields.
CONCLUSIONS: RVS has the potential to improve precision of radiation treatment delivery by detecting a significant number of setting errors. However, excessive confidence in RVS could lead to repeated errors as there is a potential for the entry of erroneous parameters into the RVS memory.

Entities:  

Mesh:

Year:  1999        PMID: 10665793     DOI: 10.1016/s0167-8140(99)00141-3

Source DB:  PubMed          Journal:  Radiother Oncol        ISSN: 0167-8140            Impact factor:   6.280


  9 in total

1.  Improving patient safety by identifying side effects from introducing bar coding in medication administration.

Authors:  Emily S Patterson; Richard I Cook; Marta L Render
Journal:  J Am Med Inform Assoc       Date:  2002 Sep-Oct       Impact factor: 4.497

2.  Implementation of electronic checklists in an oncology medical record: initial clinical experience.

Authors:  Kevin V Albuquerque; Alexis A Miller; John C Roeske
Journal:  J Oncol Pract       Date:  2011-06-17       Impact factor: 3.840

Review 3.  Organizational, technical, physical and clinical quality standards for radiotherapy.

Authors:  Marta Bogusz-Czerniewicz; Daniel Kaźmierczak
Journal:  Rep Pract Oncol Radiother       Date:  2012-06-03

4.  The report of Task Group 100 of the AAPM: Application of risk analysis methods to radiation therapy quality management.

Authors:  M Saiful Huq; Benedick A Fraass; Peter B Dunscombe; John P Gibbons; Geoffrey S Ibbott; Arno J Mundt; Sasa Mutic; Jatinder R Palta; Frank Rath; Bruce R Thomadsen; Jeffrey F Williamson; Ellen D Yorke
Journal:  Med Phys       Date:  2016-07       Impact factor: 4.071

5.  Characterization of Water-Clear Polymeric Gels for Use as Radiotherapy Bolus.

Authors:  Justus D Adamson; Tabitha Cooney; Farokh Demehri; Andrew Stalnecker; Debra Georgas; Fang-Fang Yin; John Kirkpatrick
Journal:  Technol Cancer Res Treat       Date:  2017-05-30

6.  An Investigation of Radiation Treatment Learning Opportunities in Relation to the Radiation Oncology Electronic Medical Record: A Single Institution Experience.

Authors:  Y Jessica Huang; Vikren Sarkar; Adam Paxton; Hui Zhao; Frances Fan-Chi Su; Ryan Price; Bill J Salter
Journal:  Adv Radiat Oncol       Date:  2021-09-29

7.  Clinical quality standards for radiotherapy.

Authors:  Marta Bogusz-Czerniewicz
Journal:  Contemp Oncol (Pozn)       Date:  2012-02-29

8.  Analysis of couch position tolerance limits to detect mistakes in patient setup.

Authors:  Scott W Hadley; James M Balter; Kwok L Lam
Journal:  J Appl Clin Med Phys       Date:  2009-10-29       Impact factor: 2.102

9.  Common Error Pathways in CyberKnife™ Radiation Therapy.

Authors:  Brandon T Mullins; Lukasz Mazur; Michael Dance; Ross McGurk; Eric Schreiber; Lawrence B Marks; Colette J Shen; Michael V Lawrence; Bhishamjit S Chera
Journal:  Front Oncol       Date:  2020-07-08       Impact factor: 6.244

  9 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.